Jove
Visualize
Contact Us
JoVE
x logofacebook logolinkedin logoyoutube logo
ABOUT JoVE
OverviewLeadershipBlogJoVE Help Center
AUTHORS
Publishing ProcessEditorial BoardScope & PoliciesPeer ReviewFAQSubmit
LIBRARIANS
TestimonialsSubscriptionsAccessResourcesLibrary Advisory BoardFAQ
RESEARCH
JoVE JournalMethods CollectionsJoVE Encyclopedia of ExperimentsArchive
EDUCATION
JoVE CoreJoVE BusinessJoVE Science EducationJoVE Lab ManualFaculty Resource CenterFaculty Site
Terms & Conditions of Use
Privacy Policy
Policies

Related Concept Videos

Data Validation01:15

Data Validation

206
Method validation is a crucial process in analytical chemistry designed to confirm that a given method consistently produces reliable and high-quality results. This process is essential when a method is applied to different sample matrices or when procedural modifications are made, ensuring that the results meet acceptable standards across various applications.
Key parameters for method validation include:
206
Uncertainty in Measurement: Accuracy and Precision03:37

Uncertainty in Measurement: Accuracy and Precision

74.4K
Scientists typically make repeated measurements of a quantity to ensure the quality of their findings and to evaluate both the precision and the accuracy of their results. Measurements are said to be precise if they yield very similar results when repeated in the same manner. A measurement is considered accurate if it yields a result that is very close to the true or the accepted value. Precise values agree with each other; accurate values agree with a true value. 
74.4K
Uncertainty in Measurement: Reading Instruments02:46

Uncertainty in Measurement: Reading Instruments

38.7K
Counting is the type of measurement that is free from uncertainty, provided the number of objects being counted does not change during the process. Such measurements result in exact numbers. By counting the eggs in a carton, for instance, one can determine exactly how many eggs are there in the carton. Similarly, the numbers of defined quantities are also exact. For example, 1 foot is exactly 12 inches, 1 inch is exactly 2.54 centimeters, and 1 gram is exactly 0.001 kilograms. Quantities...
38.7K
Accuracy and Precision01:52

Accuracy and Precision

10.0K
Scientists typically make repeated measurements of a quantity to ensure the quality of their findings and to evaluate both the precision and the accuracy of their results. Measurements are said to be precise if they yield very similar results when repeated in the same manner. A measurement is considered accurate if it yields a result that is very close to the true or the accepted value. Precise values agree with each other; accurate values agree with a true value.  Highly accurate...
10.0K
Voltammetry: Factors Affecting Measurements01:21

Voltammetry: Factors Affecting Measurements

193
A current produced due to the redox reactions of the analyte at the working and auxiliary electrodes is called a faradaic current. The reaction can be divided into two types. The current generated due to the reduction of the analyte is called cathodic current, and it carries a positive charge. In contrast, the current produced by analyte oxidation is known as an anodic current, and it has a negative charge. The applied potential at the working electrode determines the faradaic current flow, and...
193
Guidelines For Measuring Vital Signs01:19

Guidelines For Measuring Vital Signs

1.8K
Following these guidelines can help nurses accurately measure vital signs, assess changes in patient conditions, and provide timely treatment when necessary. Adhering closely to the guidelines ensures the accuracy and reliability of the results.
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
1.8K

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Which immunonutritional marker best predicts disease-free survival in non-metastatic colorectal Cancer? The CALLY index and CAR emerge as independent predictors from a seven-marker comparison.

Frontiers in nutrition·2026
Same author

Synovial Fluid Biomarkers Reflect Pre- and Postoperative Clinical Features in Rotator Cuff Tears, Independent of Tear Size.

Journal of orthopaedic research : official publication of the Orthopaedic Research Society·2026
Same author

Evaluation of Protective Efficacy of Formalin-Killed and Auxotrophic-Live Vibrio anguillarum Against Homologous and Heterologous Strains in Rainbow Trout (Oncorhynchus mykiss).

Journal of fish diseases·2026
Same author

Comparison of Clinical and Radiological Outcomes Between Suction Aspiration and Combination Methods of Mechanical Thrombectomy in Patients With Acute Cerebral Infarction: The COMPETE Trial.

Journal of Korean medical science·2026
Same author

Integrated Assessment of CT-Derived Fibrosis Volume and Pulmonary Function in Identifying Progressive Pulmonary Fibrosis in Connective Tissue Disease-Associated Interstitial Lung Disease (CTD-ILD).

Archivos de bronconeumologia·2026
Same author

Ambient Air Pollution and Risk of ARDS and Mortality in Moderate to Severe COVID-19.

Respirology (Carlton, Vic.)·2026
Same journal

Identification of genes associated with cervical intervertebral disc degeneration and immune cell infiltration.

Frontiers in surgery·2026
Same journal

Retrospective clinical study of endoscopic transfrontal approach vs. transSylvian-transinsular craniotomy for hypertensive intracerebral hemorrhage in basal ganglia: efficacy comparison and value of anatomical cognition of Sylvian fissure.

Frontiers in surgery·2026
Same journal

Upper lumbar disc herniation presenting as chronic right lower abdominal pain in a patient with multilevel lumbar degeneration: a case report.

Frontiers in surgery·2026
Same journal

Combined interventional and craniotomy surgery for meningioma apoplexy complicated by acute subdural hematoma: a case report and literature review.

Frontiers in surgery·2026
Same journal

Case Report: Hybrid retrograde transcarotid stenting for common carotid artery dissection secondary to acute type A aortic dissection.

Frontiers in surgery·2026
Same journal

Granulomatosis with polyangiitis initially presenting as secretory otitis media: a case report.

Frontiers in surgery·2026
See all related articles

Related Experiment Video

Updated: Aug 14, 2025

High-precision Electromagnetic Flowmeter with Empty Pipe Detection via Complex Programmable Logic Device-based Waveform Recognition
05:11

High-precision Electromagnetic Flowmeter with Empty Pipe Detection via Complex Programmable Logic Device-based Waveform Recognition

Published on: June 27, 2025

152

Vital protocols for PolyWare™ measurement reliability and accuracy.

Jong Min Lee1, Seung-Hoon Baek2, Yeon Soo Lee1

  • 1Department of BioMedical Engineering, School of BioMedical Science, Daegu Catholic University, Gyungbuk, South Korea.

Frontiers in Surgery
|January 12, 2023
PubMed
Summary
This summary is machine-generated.

Accurate polyethylene wear measurements in total hip arthroplasty (THA) require specific imaging protocols. Optimizing PolyWare™ software (PW) use involves standardized square images and centered edge detection for reliable THA wear analysis.

Keywords:
PolyWare (PW)anteversionlateral inclinationpolyethylene weartotal hip arthroplasty (THA)

More Related Videos

Continuous-Wave Propagation Channel-Sounding Measurement System - Testing, Verification, and Measurements
09:36

Continuous-Wave Propagation Channel-Sounding Measurement System - Testing, Verification, and Measurements

Published on: June 25, 2021

3.2K
Determination of Zeta Potential via Nanoparticle Translocation Velocities through a Tunable Nanopore: Using DNA-modified Particles as an Example
08:42

Determination of Zeta Potential via Nanoparticle Translocation Velocities through a Tunable Nanopore: Using DNA-modified Particles as an Example

Published on: October 26, 2016

12.4K

Related Experiment Videos

Last Updated: Aug 14, 2025

High-precision Electromagnetic Flowmeter with Empty Pipe Detection via Complex Programmable Logic Device-based Waveform Recognition
05:11

High-precision Electromagnetic Flowmeter with Empty Pipe Detection via Complex Programmable Logic Device-based Waveform Recognition

Published on: June 27, 2025

152
Continuous-Wave Propagation Channel-Sounding Measurement System - Testing, Verification, and Measurements
09:36

Continuous-Wave Propagation Channel-Sounding Measurement System - Testing, Verification, and Measurements

Published on: June 25, 2021

3.2K
Determination of Zeta Potential via Nanoparticle Translocation Velocities through a Tunable Nanopore: Using DNA-modified Particles as an Example
08:42

Determination of Zeta Potential via Nanoparticle Translocation Velocities through a Tunable Nanopore: Using DNA-modified Particles as an Example

Published on: October 26, 2016

12.4K

Area of Science:

  • Biomedical Engineering
  • Orthopedic Surgery
  • Medical Imaging Analysis

Background:

  • PolyWare™ software (PW) is the primary tool for polyethylene wear assessment in total hip arthroplasty (THA).
  • Existing PW measurements can suffer from significant inaccuracies and lack of repeatability due to imaging conditions and user subjectivity.
  • Identifying and mitigating sources of error in PW measurements is critical for reliable THA wear analysis.

Purpose of the Study:

  • To investigate the impact of various measurement conditions on PolyWare™ (PW) accuracy and reliability.
  • To establish optimal protocols for achieving precise polyethylene wear measurements in THA using PW software.

Main Methods:

  • Simulated a linear wear rate of 6.67 mm on in-vitro THA prostheses under clinical X-ray conditions.
  • Acquired X-ray images with specific acetabular lateral inclination (36.6°) and anteversion (9.0°).
  • Evaluated the influence of image resolution, edge detection area, and component positioning relative to the X-ray beam.

Main Results:

  • 1,076 × 1,076 resolution images yielded the lowest standard deviation in wear measurements (0.01 mm) and minimal blurriness.
  • Non-square edge detection areas off the femoral head center resulted in the highest degree of blurriness.
  • Eccentric placement of the femoral head (15 cm superior to the beam center) caused a maximum acetabular anteversion error of 5.3°.

Conclusions:

  • Standardized imaging protocols are essential for accurate and repeatable PolyWare™ (PW) measurements in THA.
  • Recommended protocols include using 1,076 × 1,076 square images, square edge detection areas (5x femoral head diameter) centered on the femoral head, and aligning component centers with the X-ray beam.
  • Adherence to these protocols can significantly improve the reliability of polyethylene wear quantification in THA studies.