A practical and robust method for beam blocker-based cone beam CT scatter correction.
Hehe Cui1, Xiao Jiang1, Wei Tang2
1Department of Engineering and Applied Physics, University of Science and Technology of China, Hefei, Anhui, 230026 People's Republic of China.
This study introduces an improved scatter correction method for cone beam CT (CBCT) using a beam-blocker. The new technique simplifies scatter estimation, reducing CT number errors in phantom imaging.
Area of Science:
- Medical Imaging
- Radiological Physics
- Computational Imaging
Background:
- Cone beam CT (CBCT) imaging is susceptible to scatter radiation, which degrades image quality.
- Traditional scatter correction methods using beam-blockers require tedious optimization of correction factors.
- A need exists for a more efficient and objective scatter correction method in CBCT.
Purpose of the Study:
- To develop and validate an indirect scatter estimation method for beam-blocker based CBCT scatter correction.
- To eliminate the need for optimization of correction factors in CBCT scatter correction.
- To improve the accuracy and robustness of CBCT imaging.
Main Methods:
- An indirect scatter estimation method using a beam-blocker with lead strips was developed.
- Primary signal was calculated by subtracting measured scatter, then used to estimate scatter in unblocked regions.
- Scatter signals were smoothed and applied for projection correction; CBCT was reconstructed using the Feldkamp-Davis-Kress algorithm.
Main Results:
- CT number errors were significantly reduced in phantom studies with the proposed method.
- Without a bow-tie filter, errors were reduced to 3.97±2.27 HU (Catphan, full-blocker) and 4.01±2.18 HU (head phantom, full-blocker).
- With a bow-tie filter, errors were further reduced to 2.29±1.42 HU (Catphan, full-blocker) and 2.35±1.25 HU (head phantom, full-blocker).
Conclusions:
- The proposed indirect scatter estimation method effectively corrects scatter in CBCT without tedious optimization.
- This method avoids self-influence from the beam blocker on scatter estimation.
- It offers a practical and robust solution for beam-blocker based scatter correction in CBCT.