A precise and simple isodose-volume-based verification method for HDR and LDR brachytherapy plans
H Schiefer1, S Heinze1, M Glatzer1
1Department of Radiation Oncology, Kantonsspital St.Gallen, St.Gallen, Switzerland.
Brachytherapy
|January 12, 2023
Summary
An easy isodose-based method verifies brachytherapy treatment times for high-dose-rate (HDR) and low-dose-rate (LDR) plans. This precise and simple verification method aligns with AAPM recommendations for clinical practice.
Area of Science:
- Medical Physics
- Radiation Oncology
- Radiotherapy Planning
Background:
- The American Association of Physicists in Medicine (AAPM) recommends independent treatment time calculations for brachytherapy.
- Existing dose-volume-based methods for HDR and LDR brachytherapy require specific parameters for verification.
Purpose of the Study:
- To implement and validate an easy-to-use, isodose-based verification method for HDR and LDR brachytherapy plans.
- To provide a practical alternative for independent treatment time verification in brachytherapy.
Main Methods:
- Developed an isodose-based verification method for HDR (Ir-192) and LDR (I-125 prostate) brachytherapy.
- Validated the method using 55 clinical HDR and 243 clinical LDR plans.
- Compared estimated integral dwell times/activity with calculated values based on isodose volumes.
Main Results:
- For HDR brachytherapy, mean absolute differences in integral dwell time were 0.8% ± 1.0% (single source path) and 2.7% ± 1.1% (multiple source paths).
- For LDR brachytherapy, the mean absolute difference was 1.8% ± 2.0%.
- Good agreement was achieved using the 2 Gy isodose for HDR, while LDR verification parameters depended on seed distribution.
Conclusions:
- Isodose-based verification methods are precise and simple to implement in clinical practice.
- These methods do not require specialized tools, making them accessible.
- The study recommends using an isodose at the prescribed dose for prostate HDR therapy verification.