Related Experiment Video
Updated: Aug 14, 2025

Proton Therapy Delivery and Its Clinical Application in Select Solid Tumor Malignancies
Published on: February 6, 2019
Transitioning from a COMS-based plaque brachytherapy program to using eye physics plaques and plaque simulator
Sheridan G Meltsner1, Anna Rodrigues1, Miguel A Materin2
1Department of Radiation Oncology, Duke University, Durham, North Carolina, USA.
Abstract:
The aim of this work is to describe the implementation and commissioning of a plaque brachytherapy program using Eye Physics eye plaques and Plaque Simulator treatment planning system based on the experience of one institution with an established COMS-based plaque program. Although commissioning recommendations are available in official task groups publications such as TG-129 and TG-221, we found that there was a lack of published experiences with the specific details of such a transition and the practical application of the commissioning guidelines. The specific issues addressed in this paper include discussing the lack of FDA approval of the Eye Physics plaques and Plaque Simulator treatment planning system, the commissioning of the plaques and treatment planning system including considerations of the heterogeneity corrected calculations, and the implementation of a second check using an FDA-approved treatment planning system. We have also discussed the use of rental plaques, the analysis of plans using dose histograms, and the development of a quality management program. By sharing our experiences with the commissioning of this program this document will assist other institutions with the same task and act as a supplement to the recommendations in the recently published TG-221.