Related Experiment Video
Updated: Aug 14, 2025

Post Column Derivatization Using Reaction Flow High Performance Liquid Chromatography Columns
Published on: April 26, 2016
Chemo- and enantio-selective reversed-phase HPLC analysis of rosuvastatin using a cellulose-based chiral stationary
Francesca Romana Mammone1, Paola Rotundo1, Rosella Ferretti1
1Centre for the Control and Evaluation of Medicines, Chemical Medicines Unit, Istituto Superiore di Sanità, Viale Regina Elena 299, 00161 Rome, Italy.
Abstract:
A direct reversed-phase high-performance liquid chromatographic (HPLC) method was developed for determining the content of the enantiomeric impurity of the chiral statin rosuvastatin calcium salt (RSV) in commercial tablets. The baseline enantioseparation was achieved using the Lux Cellulose-2 column and a binary linear gradient of acetonitrile and trifluoroacetic acid 0.05% in an aqueous solution. The flow rate of the mobile phases and column temperature were set at 1.0 mL min- 1 and 40 °C, respectively. In comparison with the isocratic HPLC method reported in the European Pharmacopoeia (EP) monograph for RSV, the gradient elution method offered improved chemo-and enantio-selectivity and reduced analysis times. The limits of quantitation and detection of the enantiomeric impurity were found to be 0.15 and 0.05 µg mL-1.