Published on: October 4, 2024
A novel figure of merit to investigate 68Ga PET/CT image quality based on patient weight and lesion size using
Miriam Santoro1, Giuseppe Della Gala2, Giulia Paolani1
1Department of Medical Physics, IRCCS Azienda Ospedaliero-Universitaria di Bologna, 40138 Bologna, Italy; Medical Physics Specialization School, Alma Mater Studiorum, University of Bologna, 40138 Bologna, Italy.
Introduction:
Q.Clear is a Bayesian penalised-likelihood algorithm that uses a β-value for positron emission tomography(PET)/computed tomography(CT) image reconstruction(IR). Our study proposes a novel figure of merit, named CRBV, to compare the Q.Clear performances using 68Ga PET/CT image with the ordered-subset-expectation-maximization(OSEM) algorithm and to identify the optimal β-values for these images using two phantoms mimicking normal and overweight patients.
Methods:
NEMA IQ phantom with or without a ring of water-filled plastic bags (NEMAstd and NEMAow, respectively) was acquired and reconstructed with OSEM and Q.Clear at various β-values and minutes/bed position(min/bp). Contrast recovery(CR), background variability(BV) and CRBV were calculated. Highest CRBV values were used to identify optimal β-value ranges.
Results:
Q.Clear with 250 ≤ β ≤ 800 improved CRBV compared to OSEM for all the investigated spheres and acquisition setups. Outside of this range, Q.Clear still outperformed OSEM with few exceptions depending on spheres diameters and phantoms(e.g.,β-value = 1600 for diameters ≤ 17 mm using the NEMAow phantom). Regarding the CRBV performance for IR optimization, for the 4 min/bp NEMAstd IR, β-values = 300 ÷ 350 allowed to simultaneously optimize all diameters(except for the 10 mm); for the NEMAow IR, β-values = 350 ÷ 500 were needed for diameters > 20 mm, while β-values = 200 ÷ 250 were selected for the remaining diameters. For the 2 min/bp, β-value = 500 was suitable for diameters > 17 mm in both NEMAstd and NEMAow IR, while for smaller diameters β-value = 200 and β-values = 250 ÷ 350 were obtained for NEMAstd and NEMAow, respectively.
Conclusion:
Almost all tested β-values of Q.Clear improved the CRBV compared to OSEM. In both phantoms, simulating normal and over-weight patients, optimal β-values were found according to lesion sizes and investigated acquisition times.
More Related Videos
09:49A Whole Body Dosimetry Protocol for Peptide-Receptor Radionuclide Therapy PRRT: 2D Planar Image and Hybrid 2D+3D SPECT/CT Image Methods
Published on: April 24, 2020
06:53Management of Respiratory Motion Artefacts in 18F-fluorodeoxyglucose Positron Emission Tomography using an Amplitude-Based Optimal Respiratory Gating Algorithm
Published on: July 23, 2020
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography