Endoscopic retrograde cholangiopancreatography and endoscopic ultrasound endoscope reprocessing: Variables impacting
Ashley M Ayres1, Julia Wozniak1, Jose O'Neil2
1Department of Infection Prevention and Control, UPMC Presbyterian-Shadyside, Pittsburgh, Pennsylvania.
Objective:
To evaluate variables that affect risk of contamination for endoscopic retrograde cholangiopancreatography and endoscopic ultrasound endoscopes.
Design:
Observational, quality improvement study.
Setting:
University medical center with a gastrointestinal endoscopy service performing ∼1,000 endoscopic retrograde cholangiopancreatography and ∼1,000 endoscopic ultrasound endoscope procedures annually.
Methods:
Duodenoscope and linear echoendoscope sampling (from the elevator mechanism and instrument channel) was performed from June 2020 through September 2021. Operational changes during this period included standard reprocessing with high-level disinfection with ethylene oxide gas sterilization (HLD-ETO) was switched to double high-level disinfection (dHLD) (June 16, 2020-July 15, 2020), and duodenoscopes changed to disposable tip model (March 2021). The frequency of contamination for the co-primary outcomes were characterized by calculated risk ratios.
Results:
The overall pathogenic contamination rate was 4.72% (6 of 127). Compared to duodenoscopes, linear echoendoscopes had a contamination risk ratio of 3.64 (95% confidence interval [CI], 0.69-19.1). Reprocessing using HLD-ETO was associated with a contamination risk ratio of 0.29 (95% CI, 0.06-1.54). Linear echoendoscopes undergoing dHLD had the highest risk of contamination (2 of 18, 11.1%), and duodenoscopes undergoing HLD-ETO and the lowest risk of contamination (0 of 53, 0%). Duodenoscopes with a disposable tip had a 0% contamination rate (0 of 27).
Conclusions:
We did not detect a significant reduction in endoscope contamination using HLD-ETO versus dHLD reprocessing. Linear echoendoscopes have a risk of contamination similar to that of duodenoscopes. Disposable tips may reduce the risk of duodenoscope contamination.
Related Concept Videos
Endoscopic Procedures V: ERCP
Patient...
Cleaning, Sterilization, and Disinfection
Cleaning
The cleaning process usually involves using water with detergents or enzymatic cleaner and removing foreign material from objects and surfaces, including organic material such as body fluids or inorganic material like soil. Cleaning is performed before high-level disinfection and sterilization because foreign materials on the cover of the devices interfere with process...
Endoscopic Procedures I: Esophagogastroduodenoscopy
During an EGD, the endoscope can be used to:
Standard Precaution
Hand hygiene is the most crucial means to prevent the transmission of disease. Employers are legally required to provide their workers with personal protective equipment (PPE) to minimize exposure or contact with...
Endoscopic Procedures IV: Sigmoidoscopy and Laproscopy
Sigmoidoscopy
Sigmoidoscopy is a diagnostic procedure that uses a flexible sigmoidoscope equipped with a light source and camera to examine the rectum and sigmoid colon. The procedure involves inserting the tube through the anus...
Endoscopic Studies II: Thoracocentesis
Description
Excess pleural fluid or air may accumulate in some respiratory disorders in the thoracic cavity. To treat pleural effusion, a physician conducts thoracentesis by carefully piercing the chest wall and entering...