Effectiveness of radiation protection systems in the cardiac catheterization laboratory: a comparative study
Victoria L Cammann1,2, Victor Schweiger1,2, Maciej Cieslik1
1Department of Cardiology, University Hospital Zurich, University Heart Center, Raemistrasse 100, 8091, Zurich, Switzerland.
Background:
As numbers and complexity of percutaneous coronary interventions are constantly increasing, optimal radiation protection is required to ensure operator safety. Suspended radiation protection systems (SRPS) and protective scatter-radiation absorbing drapes (PAD) are novel methods to mitigate fluoroscopic scattered radiation exposure. The aim of the study was to investigate the effectiveness regarding radiation protection of a SRPS and a PAD in comparison with conventional protection.
Methods:
A total of 229 cardiac catheterization procedures with SRPS (N = 73), PAD (N = 82) and standard radiation protection (N = 74) were prospectively included. Real-time dosimeter data were collected from the first operator and the assistant. Endpoints were the cumulative operator exposure relative to the dose area product [standardized operator exposure (SOE)] for the first operator and the assistant.
Results:
For the first operator, the SRPS and the PAD significantly decreased the overall SOE compared to conventional shielding by 93.9% and 66.4%, respectively (P < 0.001). The protective effect of the SRPS was significantly higher compared to the PAD (P < 0.001). For the assistant, the SRPS and the PAD provided a not statistically significant reduction compared to conventional shielding in the overall SOE by 38.0% and 30.6%, respectively.
Conclusions:
The SRPS and the PAD enhance radiation protection significantly compared to conventional protection. In most clinical scenarios, the protective effect of SRPS is significantly higher than the additional protection provided by the PAD. Comparison of the additional radiation protection provided by protective scatter-radiation absorbing drapes (PAD) and the suspended radiation protection system (SRPS) system over standard protection with lead aprons.
More Related Videos
Related Concept Videos
Cardiac Catheterization I: Pre-Procedure Overview
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Imaging Studies for Cardiovascular System V: CT
Cardiac Catheterization III: Left Heart Catheterization
Cardiac Catheterization IV: Nursing Management
Cardiac Catheterization II: Right Heart Catheterization