Characterization of Ethos therapy systems for adaptive radiation therapy: A multi-machine comparison
Agustinus J van de Schoot1, Daan Hoffmans2, Karel M van Ingen1
1Department of Radiation Oncology, Amsterdam University Medical Center - location University of Amsterdam, Amsterdam, the Netherlands.
This study shows that multiple Ethos radiation therapy systems perform consistently, with minimal variations in dosimetric and mechanical characteristics. This ensures reliable patient treatment across different machines.
Area of Science:
- Medical Physics
- Radiation Oncology
Background:
- The Ethos therapy system enables online CBCT-guided adaptive radiation therapy.
- Clinical implementation of multiple systems necessitates machine characterization and variation quantification for patient interchangeability.
- Limited vendor-independent data exists on Ethos machine performance.
Purpose of the Study:
- To characterize dosimetric and mechanical features of Ethos machines.
- To quantify variations between individual Ethos machines.
- To provide vendor-independent performance data.
Main Methods:
- Six Ethos machines with 6-MV FFF beams were characterized.
- Point doses, profiles, and dose distributions were measured and compared to vendor data and calculations.
- Key parameters quantified include leaf transmission, DLG, couch attenuation, isocenter accuracy, and kV-MV coincidence.
- 1D and 3D gamma analysis was used to assess agreement.
Main Results:
- Excellent agreement was observed between measurements and vendor data/dose calculations (mean γ1%/1mm < 0.30).
- All machines demonstrated consistent performance in profiles, point measurements, and dose distributions.
- Minimal machine variations were found for DLG (0.05 mm), leaf transmission (0.001%), isocenter accuracy (0.08 mm), kV-MV coincidence (0.15 mm), couch attenuation (0.69%), and CBCT dose (0.29 mGy).
Conclusions:
- Ethos therapy systems exhibit excellent agreement with vendor data and pre-configured beam models.
- High consistency was demonstrated across all tested machines.
- These findings establish benchmark data for the clinical introduction of Ethos systems.
Related Concept Videos
Humanistic Therapy
Rational Emotive Behavior Therapy
Psychodynamic Therapy
Interpersonal Psychotherapy
Family Therapy
Strategic Family Therapy
Strategic family therapy emphasizes resolving communication barriers and improving problem-solving abilities...
Modeling in Therapy
Participant Modeling
Participant modeling involves therapists demonstrating calm and effective behaviors in...