Generation of Electronic Cigarette Aerosol by a Third-Generation Machine-Vaping Device: Application to Toxicological Studies
Published on: August 25, 2018
Analysis of vaporized caffeine in smoke from e-cigarettes using liquid chromatography-tandem mass spectrometry and
Makoto Takada1, Suzuna Saruwatari1, Yutaro Yanagita1
1Faculty of Pharmaceutical Sciences, Sanyo-Onoda City University, 1-1-1 Daigakudori, Sanyo-Onoda, Yamaguchi, 756-0884, Japan.
Purpose:
Electronic cigarettes (e-cigarettes) are used widely, and e-cigarettes containing caffeine (Caf) have recently become commercially available. However, no risk evaluation of these Caf-containing products has been performed to date. Such an evaluation requires a sensitive analytical method for quantifying Caf in smoke from e-cigarettes. The aim of this study was to establish a liquid chromatography-tandem mass spectrometry (LC-MS/MS) method for quantifying vaporized Caf from commercially available e-cigarettes, and to determine minor components related to Caf in cigarette smoke extract (CSE).
Methods:
A sampling system for Caf using a suction pump was designed and sampling conditions were optimized.
Results:
The optimized LC-MS/MS conditions allowed the sensitive determination of Caf in smoke with a limit of detection of 0.03 ng/mL at a signal-to-noise ratio of 3. The method was applied to CSEs from five e-cigarette products and the concentration of Caf ranged from 0.894 ± 0.090 to 3.32 ± 0.14 μg/mL smoke (n = 3). Additionally, minor components related to Caf, such as theobromine, theophylline, and paraxanthine, were detected in CSE and in e-liquid at very low concentrations, indicating that they were impurities in e-liquid and vaporized along with Caf.
Conclusion:
This is the first report to determine the concentration of vaporized Caf using an LC-MS/MS method and to clarify several minor components in smoke from e-cigarettes.
More Related Videos
09:30A Microcontroller Operated Device for the Generation of Liquid Extracts from Conventional Cigarette Smoke and Electronic Cigarette Aerosol
Published on: January 18, 2018
10:44Comparing the Effects of Electronic Cigarette Vapor and Cigarette Smoke in a Novel In Vivo Exposure System
Published on: May 24, 2017
Related Concept Videos
Electrospray Ionization (ESI) Mass Spectrometry
ESI utilizes electrical energy to transfer ions from the liquid phase of the sample into the...
Mass Spectrometry: Complex Analysis
GC–MS is a powerful hyphenated method commonly used in forensics and environmental...
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall....
High-Performance Liquid Chromatography: Types of Detectors
Volatilization