Related Experiment Video
Updated: Aug 16, 2026

Creating Rigidly Stabilized Fractures for Assessing Intramembranous Ossification, Distraction Osteogenesis, or Healing of Critical Sized Defects
Published on: April 11, 2012
Displaced olecranon fractures in adults. Biomechanical analysis of fixation methods
D F Murphy1, W B Greene, J A Gilbert
1Division of Orthopaedic Surgery, University of North Carolina School of Medicine, Chapel Hill 27514.
Abstract:
Four commonly used methods of internal fixation for displaced olecranon fractures were analyzed in a biomechanical model to determine strength of fixation. A transverse osteotomy in fresh cadaver specimens was internally fixed and was tested to failure by rapid loading. Screw plus wire in combination provided the greatest strength of fixation. Energy to failure averaged 31.0 Nm for cancellous screw, 32.0 Nm for figure-of-eight wire, 35.3 Nm for Arbeitsgemeinschaft fur Osteosynthesefragen (AO) tension band, and 39.4 Nm for screw-plus-wire combination. Comparison of screw and wire in combination to screw alone or to figure-of-eight wire demonstrated a significant difference (p less than 0.05). The difference between screw plus wire and AO tension-band fixation did not achieve statistical significance (0.2 greater than p greater than 0.1).
More Related Videos
15:11Surgical Fixation of Sternal Fractures: Preoperative Planning and a Safe Surgical Technique Using Locked Titanium Plates and Depth Limited Drilling
Published on: January 5, 2015
11:21Development of a Novel Internal Fixation Model for Rat Radial Fractures: Fracture Healing Assessment and Dorsal Root Ganglion Isolation
Published on: March 13, 2026