Daniel Lewin1, Gaik Nersesian2, Pia Lanmüller2
1Department of Cardiothoracic and Vascular Surgery, German Heart Center Berlin, Berlin, Germany.
Background:
Impella 5.0 and 5.5 (summarized as Impella 5+) are microaxial, catheter-based left ventricular assist devices (LVAD) that are implanted via a vascular graft sutured to the axillary artery and provide blood flow of up to 5.5 liter/min. This study aims to investigate the incidence of long-term complications following circulatory support with Impella 5+.
Methods:
A single-center retrospective analysis of 203 consecutive adult patients treated between January 2017 and September 2021 with a surgically implanted Impella 5.0 or 5.5 via a vascular graft sutured to the axillary artery.
Results:
The median Impella support duration was 8 days. Of 203 patients, 78 (38.4%) died while on temporary mechanical circulatory support. Fifty-five (27.1%) were successfully weaned from Impella 5+ and 70 (34.5%) were bridged to a durable LVAD with a median follow-up time of 232 (IQR 68.5, 597) days after Impella 5+ explantation. In 119 of these patients, the Impella was explanted and the vascular graft was shortened, ligated, and pushed under the pectoralis muscle; in 6 patients early graft infection prompted complete graft removal during explantation. In addition, 13 patients (10.9%) developed a late-onset graft infection after a median of 86 days, requiring complete (n = 10) or partial (n = 2) explantation of the retained graft. In 1 patient, the graft infection was successfully treated by conservative therapy. Our analysis identified no specific risk factors for graft infections. Of the 203 patients, 5 (2.5%) developed a brachial plexus injury resulting in neurological dysfunction.
Conclusions:
In 10.9% of patients, retaining the vascular graft was complicated by a late graft infection. Complete explantation of the graft prosthesis may decrease the infection rate, but may in turn increase the risk of brachial plexus injury. On the other hand, this method offers the possibility of bedside explantation.