Related Experiment Video
Updated: Aug 13, 2025

X-ray Dose Reduction through Adaptive Exposure in Fluoroscopic Imaging
Published on: September 11, 2011
ABSORBED DOSE AND IMAGE QUALITY ACCORDING TO THE SCAN MODE AND SHIELD POSITION IN PEDIATRIC BRAIN COMPUTED TOMOGRAPHY
Eunmi Bae1, Youngkwon Cho2, Sungchul Kim3
1Department of Diagnostic Radiology, Korea Medical Institute Suwon Center, Gwonseon-gu, Suwon-si, Gyeonggi-do, Korea.
Abstract:
In this study, computed tomography dose index (CTDIvol), dose-length product (DLP), organ dose and signal-to-noise ratio (SNR) were measured in pediatric brain helical and volumetric CT scans using a pediatric phantom (equivalent to a 10-year-old) and optically stimulated luminescence dosemeters (OSLD) as the scan protocol used in Hospital A. Volumetric scans showed lower CTDIvol, DLP and organ dose than helical scans, and the SNR showed similar results. The organ doses were lower in the phantoms protected with front, rear or front and rear shielding than in those without shielding. However, no statistically significant differences were found among the different shield locations. In pediatric brain CT scans, to reduce exposure dose and patient discomfort while maintaining image quality, a volumetric scan rather than a helical scan was preferred, and a good shielding effect was observed with front or rear shielding.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
X-ray Imaging