Related Experiment Video
Updated: Aug 13, 2025

06:54
Clinical-oriented Three-dimensional Gait Analysis Method for Evaluating Gait Disorder
Published on: March 4, 2018
14.2K
Reliability and applicability of a low-cost, camera-based gait evaluation method for clinical use
Savvas Spanos1,2, Asimakis Kanellopoulos2,3, Kyriakos Petropoulakos1
1Human Performance and Rehabilitation Laboratory, Department of Physiotherapy, University of Thessaly, Lamia, Greece.
Expert Review of Medical Devices
|January 20, 2023
Summary
Kinovea® software offers a reliable and user-friendly method for clinical gait analysis. This study confirms its objectivity and low cost, with good to excellent reliability for most parameters, excluding pelvis position.
Area of Science:
- Biomechanics
- Movement Analysis
- Clinical Assessment
Background:
- Need for reliable and accessible clinical gait assessment tools.
- Kinovea® is an open-access software for human movement analysis.
- Limited research on Kinovea®'s repeatability in gait analysis.
Purpose of the Study:
- Evaluate Kinovea® software for objective, quantitative, low-cost, and user-friendly gait evaluation.
- Assess the applicability and reliability of Kinovea® in a clinical setting.
Main Methods:
- Collected gait data from 44 healthy subjects using two smartphones.
- Recorded gait in sagittal and frontal planes.
- Utilized Kinovea® software to compute kinematical and spatio-temporal gait parameters.
Main Results:
- Demonstrated good to excellent intra- and inter-rater reliability for video processing and measurement procedures.
- Identified minimal random measurement errors for most parameters.
- Noted no significant random measurement error, except for pelvis position calculation.
Conclusions:
- Kinovea® software provides an objective, quantitative, reliable, and cost-effective gait evaluation method for clinical use.
- The software is easy to use in clinical settings.
- Accuracy for pelvis position calculation requires further attention.