Non-Steroidal Drug Interferences in a Quantitative Multisteroid LC-MS/MS Assay
Valentin Braun1,2, Hermann Stuppner1, Christoph Seger1,2
1Institute of Pharmacy/Pharmacognosy, CCB-Centrum of Chemistry and Biomedicine, University of Innsbruck, Innrain 80-82, A-6020 Innsbruck, Austria.
Abstract:
Screening for possible interferences from steroidal compounds other than the target analytes (endogenous or exogenous) is well established in LC-MS/MS assay development for steroid quantification in a routine clinical setting. However, interferences from non-steroidal substances have, hitherto, not been explored. After screening more than 150 pharmaceuticals and their metabolites by analyzing commercial quality control samples from TDM analysis kits (Recipe, Chromsystems) with a multisteroid LC-MS/MS assay (protein precipitation followed by HybridSPE filtration, biphenyl column, methanol-water gradient with NH4F additive), we can report the finding of two newly discovered potential interferences from non-steroidal drugs. Antidepressant paroxetine (PX) was identified as an interference to 17-hydroxyprogesterone (17P), and α-hydroxytriazolam (α-OH-TZM)-a major metabolite of benzodiazepine triazolam (TZM)-was identified as an interference to aldosterone (ALDO). Despite different elemental and structural compositions and nominal masses, the M+1 isotopologues of PX and α-OH-TZM produced overlapping signals in ion traces monitored for the respective analytes (m/z 331 → 109/97 and 361→315/343, respectively). PX and TZM are frequently prescribed drugs, and their therapeutic ranges are far exceeding the reference ranges of 17P or ALDO (µmol vs nmol); therefore, these interferences should be considered clinically relevant. Striving for faster multi-analyte methods with high sample turnover, especially in the field of steroid quantification, can limit assay selectivity and specificity. Therefore, supported by the findings of this study, screening for potential interferences in multi-analyte LC-MS/MS method development should not cover only substances of the same class but also include a set of common drugs.
More Related Videos
10:13Author Spotlight: An Efficient Methodology to Confidently Differentiate and Characterize Fentanyl Analogs
Published on: November 8, 2024
10:37Deep Proteome Profiling by Isobaric Labeling, Extensive Liquid Chromatography, Mass Spectrometry, and Software-assisted Quantification
Published on: November 15, 2017
Related Concept Videos
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences
Atomic Absorption Spectroscopy: Interference
Spectral interference occurs when signals from other elements or molecules overlap with the analyte signal, falsely elevating or masking the analyte's absorbance. This interference can be corrected using Zeeman,...