Using an Automated Hirschberg Test App to Evaluate Ocular Alignment
Published on: March 24, 2020
The blinq™ Vision Screener in Detection of Amblyopia and Strabismus
Savannah Reaves Monahan1, Samantha M Smith1, Elizabeth Thompson Gammeltoft1
1From the Storm Eye Institute, Medical University of South Carolina, Charleston, South Carolina.
Purpose:
The blinq (Rebion Inc) is a new screening device designed to directly detect amblyopia and strabismus rather than amblyopia risk factors. We performed an independent assessment of the effectiveness of the blinq in detecting amblyopia and strabismus.
Design:
Prospective clinical validity analysis of a screening device based on sensitivity and specificity.
Methods:
Children presenting for examination in the pediatric ophthalmology clinic underwent screening with the blinq before examination by a pediatric ophthalmologist blinded to the screening results. Results of the blinq and examination findings of strabismus or amblyopia were compared.
Results:
In our cohort of 267 children with an average age of 6.3 years, the sensitivity of the blinq to detect amblyopia or any constant strabismus was 87.5% (78.2%-93.8%) and specificity was 51.3% (43.9%-58.7%). Using the previously described "appropriate referral gold standard" criteria, including children with intermittent strabismus and high refractive error, the sensitivity increased to 91.3% and the specificity to 63.2%. We found a high number of children (44 [16%]) upon whom the blinq timed out and were included as automatic referrals.
Conclusions:
Our results support use of the blinq as a screening device to detect amblyopia and strabismus in children.