Background:
SARS-CoV-2 rapid antigen tests (RATs) are in high demand for reducing the spread of SARS-CoV-2. Reduced involvement from health care professionals (HCPs) for collection and interpretation could significantly foster the wide-spread implementation of RATs, but data evaluating RATs, when used by lay people, is limited.
Objective:
To valuate agreement between BD Veritor test results for self- and HCP-collected specimens, and visually- and analyzer-interpreted results.
Methods:
Individuals with onset of COVID-19 symptoms within five days of enrollment had three nasal swabs collected; one self-collected and the other two HCP-collected. One HCP-collected swab was stored for future testing while the order of the other two (self and HCP) was randomized before testing. with the BD Veritor System for Rapid Detection of SARS-CoV-2. Results were first assessed visually, followed by interpretation with the analyzer.
Results:
When self-collection was compared to HCP collection for SARS-CoV-2 detection, interpretation by analyzer resulted in positive percent agreement (PPA) of 94.7% (95% CI 82.7, 98.5) and negative percent agreement (NPA) of 99.0% (95% CI 97.5, 99.6). When visual interpretation was compared to analyzer-read results, collection by HCPs had a PPA of 97.4% (95% CI 86.5, 99.5) and NPA of 99.8% (95% CI 98.6, 100.0) while self-collection resulted in PPA of 94.9% (95% CI 83.1, 98.6) and NPA of 99.8% (95% CI 98.6, 100).
Conclusions:
Similar PPA and NPA were observed for self- and HCP-collected specimens as well as visually- and analyzer-interpreted tests. The equivalence in performance supports the use of expanded collection and testing methods.