Analytical and clinical performances of seven direct detection assays for SARS-CoV-2
Yasufumi Matsumura1, Wataru Yamazaki2, Taro Noguchi1
1Department of Clinical Laboratory Medicine, Kyoto University Graduate School of Medicine, 54 Shogoin-kawaharacho, Sakyo-ku, Kyoto, 6068507, Japan.
Background:
Direct detection tests for severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) that bypass complicated nucleic acid/antigen purification steps are promising tools for the rapid diagnosis of coronavirus disease 2019 (COVID-19).
Methods:
To determine the analytical and clinical diagnostic performances of the direct detection assays, we compared 6 direct molecular detection assays, including two loop-mediated isothermal amplification (LAMP) assays and one lateral flow antigen assay, against the reference extraction-based RT-PCR assay using 183 respiratory samples (87 nasopharyngeal swabs, 51 saliva samples, and 45 sputum samples).
Results:
Analytical sensitivity analysis showed that the direct RT-PCR assay of Toyobo exhibited the lowest LOD of 1,000 copies/mL. Compared with the 80 positive and 103 negative samples based on the reference assay, the Toyobo assay had the highest positive percent agreement (PPA) of 96.3%, followed by the two direct RT-PCR assays of Takara and Shimadzu and one LAMP assay of Eiken (86.3-87.5%). The Fujirebio antigen assay had the lowest PPA of 44.7% among the assays tested. The negative percent agreement of these direct detection assays was 100%, except for the Eiken assay (96.3%).
Conclusions:
Large differences in PPA existed among the direct detection tests. Laboratories need to take these characteristics into consideration before implementing these assays.