As low as diagnostically acceptable dose imaging in maxillofacial trauma: a reference quality approach
Gerlig Widmann1, Hannes Schönthaler1, Alexander Tartarotti2
1Department of Radiology, Medical University of Innsbruck, Innsbruck, Austria.
Dento Maxillo Facial Radiology
|January 23, 2023
Summary
This study demonstrates that reference quality metrics from clinical As-low-as-diagnostically-acceptable (ALADA) dose protocols can effectively guide the selection of new ALADA protocols for CT scanners, ensuring diagnostic quality at reduced radiation doses.
Area of Science:
- Medical Imaging
- Radiology
- Radiation Dose Management
Background:
- Current diagnostic reference levels for CT scans are often higher than As-low-as-diagnostically-acceptable (ALADA) doses.
- Improving dose management in CT imaging is crucial for patient safety.
Purpose of the Study:
- To test a reference quality approach for selecting potential ALADA dose protocols for various CT scanner models.
- To benchmark new ALADA protocols against established ones using phantom and cadaver studies.
Main Methods:
- Phantom studies evaluated spatial resolution, contrast resolution, CNR, noise, and sharpness for a reference ALADA protocol and various test protocols across different CT scanners and parameters.
- Test protocols were selected based on non-inferior quality metrics compared to the reference protocol.
- Selected low-dose protocols were verified in a cadaver study for diagnostic accuracy in detecting maxillofacial fractures and CNR of orbital structures.
Main Results:
- Phantom quality metrics identified potential ALADA dose protocols with doses as low as 1.73-1.92 mGy.
- Cadaver studies confirmed non-inferiority of selected low-dose protocols in detecting maxillofacial fractures and achieving adequate CNR for orbital soft tissues compared to reference images.
Conclusions:
- Reference quality metrics derived from clinical ALADA dose protocols are a viable method for selecting potential ALADA protocols.
- This approach can guide the development of lower-dose CT protocols across different scanner models while maintaining diagnostic image quality.