Related Experiment Video
Updated: Aug 13, 2025

Minimally Invasive Treatment for Thoracolumbar Burst Fracture Using Sagittal Alignment Screws and A Trauma Reduction Device
Published on: November 8, 2024
Fracture Gap Closure and Reduction Are Affected by the Orientation of the Headless Compression Screw
John P Mickley1, Daniel J Lynch1, Adam M Gordon2
1The Ohio State University College of Medicine, Columbus, USA.
Background:
We evaluated the impact of a variable-pitch headless screw's angle of insertion relative to the fracture plane on fracture gap closure and reduction.
Methods:
Variable-pitch, fully threaded headless screws were inserted into polyurethane blocks of "normal" bone model density using a custom jig. Separate trials were completed with a 28-mm screw placed perpendicular and oblique/longitudinal to varying fracture planes (0°, 15°, 30°, 45°, and 60°). Fluoroscopic images were taken after each turn during screw insertion and analyzed. Initial screw push-off, residual fracture gap at optimal fracture gap reduction, and malreduction were determined in each trial. Statistical analysis was performed via a 1-way analysis of variance followed by Student t tests.
Results:
Malreduction was found to be significantly different between the perpendicular (1.88 mm ± 1.38) and the oblique/longitudinal (0.58 mm ± 0.23) screws. The malreduction increased for the perpendicular screw as the fracture angle increased (60° > 45°=30° > 15° > 0°). Residual fracture gap at optimal fracture gap reduction was also found to be significantly different between the perpendicular (0.97 ± 0.42) and oblique/longitudinal (1.43 ± 1.14) screws. The residual fracture gap increased for the oblique/longitudinal screw as the fracture angle increased, although the oblique/longitudinal screw with a 60° fracture angle was the only configuration significantly larger than all the other configurations. Screw push-off was not found to be significantly different between the oblique/longitudinal screw and perpendicular screw trials.
Conclusions:
The perpendicular screw had a larger malreduction that increased with fracture angle, whereas the oblique/longitudinal screw had a larger residual fracture gap that increased with fracture angle.
More Related Videos
Related Concept Videos
Fractures: Bone Repair
Minor fractures with no bone displacement are treated by immobilizing the fractured bone using a cast or splint. However, in the case of fractures with displaced bones, the broken bones are repositioned before immobilization to ensure successful healing without deformation and loss of function. The realignment of fractured bone ends is performed through a process called reduction. If the...
Frictional Forces on Screws
A jack with a square-threaded screw is a mechanical device used to lift heavy loads by applying a force at its handle. When the force is applied, the screw turns, raising the load. The screw can...
Behavior of Concrete Under Compressive Load
As the concrete specimen fractures under...
Self-Locking Screw
A square-threaded screw jack carrying a load is considered self-locking if the screw retains its position even after the moment applied to it is removed.