New perforated radiation shield for anesthesiologists: Monte Carlo simulation of effects
Ayaka Yanagawa1, Takeshi Takata2, Taichi Onimaru1
1Department of Anesthesia, Teikyo University, Tokyo 173-8605, Japan.
Abstract:
Catheterization for structural heart disease (SHD) requires fluoroscopic guidance, which exposes health care professionals to radiation exposure risk. Nevertheless, existing freestanding radiation shields for anesthesiologists are typically simple, uncomfortable rectangles. Therefore, we devised a new perforated radiation shield that allows anesthesiologists and echocardiographers to access a patient through its apertures during SHD catheterization. No report of the relevant literature has described the degree to which the anesthesiologist's radiation dose can be reduced by installing radiation shields. For estimating whole-body doses to anesthesiologists and air dose distributions in the operating room, we used a Monte Carlo system for a rapid dose-estimation system used with interventional radiology. The simulations were performed under four conditions: no radiation shield, large apertures, small apertures and without apertures. With small apertures, the doses to the lens, waist and neck surfaces were found to be comparable to those of a protective plate without an aperture, indicating that our new radiation shield copes with radiation protection and work efficiency. To simulate the air-absorbed dose distribution, results indicated that a fan-shaped area of the dose rate decrease was generated in the area behind the shield, as seen from the tube sphere. For the aperture, radiation was found to wrap around the backside of the shield, even at a height that did not match the aperture height. The data presented herein are expected to be of interest to all anesthesiologists who might be involved in SHD catheterization. The data are also expected to enhance their understanding of radiation exposure protection.
More Related Videos
11:40A Co-culture Method to Investigate the Crosstalk Between X-ray Irradiated Caco-2 Cells and PBMC
Published on: January 30, 2018
09:03Radiotracer Administration for High Temporal Resolution Positron Emission Tomography of the Human Brain: Application to FDG-fPET
Published on: October 22, 2019
Related Concept Videos
Radiation: Applications
The average...
X-ray Imaging
Absorption of Radiation
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...