Surface and mechanical properties of polyurethane central venous catheters after repeated contact with chemotherapy
Afif Khzam1, Johanna Saunier1, Léa Carpentier2
1Matériaux et Santé, UFR de pharmacie, Université Paris-Saclay, Orsay, France.
Abstract:
This article investigates the impact of the interactions between polyurethane central venous catheters and solutions containing excipients used in cisplatin and paclitaxel formulations. Changes to the properties of catheters and the leaching of catheter additives into the infused solutions were studied while these solutions were infused cyclically for several months. Chemotherapy treatment was mimicked in vitro in compliance with hospital practices. The treatment cycle was repeated 10 times, using solutions containing only the excipients. After 10 treatment cycles, no physical or chemical degradation of the catheter was observed. Mechanical performances were stable, but surface modifications occurred, causing the surface to become more hydrophobic. A loss in polyurethane antioxidant amount was observed in part due to a leaching phenomenon.