Safe and effective delivery of high-power, short-duration radiofrequency ablation lesions with a flexible-tip
Leon M Ptaszek1, Jacob Koruth2, Pasquale Santangeli3
1Cardiac Arrhythmia Service, MGH Heart Center, Massachusetts General Hospital, Harvard Medical School, Boston, Massachusetts.
Background:
High-power, short-duration (HPSD) radiofrequency ablation (RFA) may reduce ablation time. Concerns that catheter-mounted thermocouples (TCs) can underestimate tissue temperature, resulting in elevated risk of steam pop formation, potentially limit widespread adoption of HPSD ablation.
Objective:
The purpose of this study was to compare the safety and efficacy of HPSD and low-power, long-duration (LPLD) RFA in the context of pulmonary vein isolation (PVI).
Methods:
An open-irrigated ablation catheter with a contact force sensor and a flexible-tip electrode containing a TC at its distal end (TactiFlexTM Ablation Catheter, Sensor EnabledTM, Abbott) was used to isolate the left pulmonary veins (PVs) in 12 canines with HPSD RFA (50 W for 10 seconds) and LPLD RFA (30 W for a maximum of 60 seconds). PVI was assessed at 30 minutes and 28 ± 3 days postablation. Computed tomographic scans were performed to assess PV stenosis after RFA. Lesions were evaluated with histopathology.
Results:
A total of 545 ablations were delivered: 252 with LPLD (0 steam pops) and 293 with HPSD RFA (2 steam pops) (P = .501). Ablation time required to achieve PVI was >3-fold shorter for HPSD than for LPLD RFA (P = .001). All 24 PVs were isolated 30 minutes after ablation, with 12/12 LPLD-ablated and 11/12 HPSD-ablated PVs still isolated at follow-up. Histopathology revealed transmural ablations for HPSD and LPLD RFA. No major adverse events occurred.
Conclusion:
An investigational ablation catheter effectively delivered RFA lesions. Ablation time required to achieve PVI with HPSD with this catheter was >3-fold shorter than with LPLD RFA.