Long-term stability of over-the-counter cuffless blood pressure monitors: a proposal
Toshiyo Tamura1, Shigeru Shimizu2, Nobuhiro Nishimura2
1Future Robotics Organization, Waseda University, Tokyo, Japan.
Insights
New criteria for cuffless blood pressure (CL-BP) monitors are proposed to ensure accuracy and long-term stability for at-home hypertension management. These standards aim to guide regulatory approval for over-the-counter devices.
Area of Science:
- Biomedical Engineering
- Cardiovascular Technology
- Medical Device Regulation
Background:
- Blood pressure monitoring is crucial for cardiovascular health.
- Cuff-based sphygmomanometers are standard but cumbersome.
- Emerging cuffless blood pressure (CL-BP) monitors require regulatory oversight for home use.
Purpose of the Study:
- To propose simple criteria for over-the-counter (OTC) CL-BP monitoring.
- To establish standards for accuracy and long-term stability in CL-BP devices.
- To facilitate regulatory approval for consumer-grade CL-BP monitors.
Main Methods:
- Reviewed history of sphygmomanometry and current protocols.
- Defined key CL-BP monitoring components: resting accuracy, dynamic accuracy, and long-term stability.
- Proposed intermittent measurements and a new experimental protocol for stability assessment.
- Validated against an approved automated sphygmomanometer.
Main Results:
- The proposed criteria address resting accuracy, dynamic changes, and long-term stability.
- Intermittent measurements ensure active accuracy matches resting accuracy.
- The tested device demonstrated long-term stability within acceptable propagation error margins.
- Static accuracy: Systolic -0.12 ± 5.49 mmHg, Diastolic -1.17 ± 5.06 mmHg.
- Long-term stability: Systolic -3.38 ± 7.1 mmHg, Diastolic -1.38 ± 5.4 mmHg.
Conclusions:
- The proposed criteria and experimental protocol are promising for evaluating CL-BP devices.
- Further research is needed to establish comprehensive manufacturing standards.
- Ensuring high-quality evidence is essential for regulatory approval of OTC CL-BP monitors.
Abstract:
Blood pressure is an important cardiovascular parameter. Currently, the cuff-based sphygmomanometer is a popular, reliable, measurement method, but blood pressure monitors without cuffs have become popular and are now available without a prescription. Blood pressure monitors must be approved by regulatory authorities. Current cuffless blood pressure (CL-BP) monitors are not suitable for at-home management and prevention of hypertension. This paper proposes simple criteria for over-the-counter CL-BP monitoring. First, the history of the sphygmomanometer and current standard blood pressure protocol are reviewed. The main components of CL-BP monitoring are accuracy during the resting condition, accuracy during dynamic blood pressure changes, and long-term stability. In this proposal we recommend intermittent measurement to ensure that active measurement accuracy mirrors resting condition accuracy. A new experimental protocol is proposed to maintain long-term stability. A medically approved automated sphygmomanometer was used as the standard device in this study. The long-term accuracy of the test device is based on the definition of propagation error, i.e., for an oscillometric automated sphygmomanometer (5 ± 8 mmHg) ± the error for the test device static accuracy (-0.12 ± 5.49 mmHg for systolic blood pressure and - 1.17 ± 5.06 mmHg for diastolic blood pressure). Thus, the long-term stabilities were - 3.38 ± 7.1 mmHg and - 1.38 ± 5.4 mmHg, which satisfied propagation error. Further research and discussion are necessary to create standards for use by manufacturers; such standards should be readily evaluated and ensure high-quality evidence.
Supplementary Information:
The online version contains supplementary material available at 10.1007/s12553-023-00726-6.
Related Concept Videos
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Errors occurring during blood pressure monitoring
Several factors...
Assessment of blood pressure in brachial artery(two-step method)
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Holter Monitor: 24-Hour Monitoring