Background:
The use of high flow oxygen therapy (HFOT) has significantly escalated during the COVID-19 pandemic. HFOT can be delivered through both dedicated devices and ICU ventilators. HFOT can be administered to a patient via a nasal cannula (NC). In intubated patients, a tracheal cannula (TC) is used instead. In this study, we aim to compare the work of breathing (WOB) using a TC or NC and to explore whether differences exist among HFOT devices.
Methods:
Seven HFOT devices (three dedicated and four ICU ventilators) were connected to a manikin head (Laerdal Medical) through a NC (Optiflow 3S, large size, Fisher and Paykel Healthcare) or a TC (OPT 970 Optiflow+, Fisher and Paykel Healthcare). Each device was also attached to a manikin head that was connected to a lung simulator (ASL5000, Ingmar Medical), set at 40 ml/cmH2O compliance, 10 cmH2O/L/s resistance, and sinusoidal inspiratory effort (muscular pressure 10 cmH2O, rate 30 breaths/min). HFOT was delivered at 40 L/min and at 21% inspired oxygen fraction. The total WOB per breath and its resistive and elastic components were automatically analyzed breath by breath over the last 20 breaths by using Campbell's diagram.
Results:
The WOB and its resistive and elastic components were significantly lower with the TC than with the NC for every device, and systematically lower with the reference device than with others. These differences were, however, very small and may be not clinically relevant.
Conclusion:
The WOB is lower with the TC than with the NC and with the reference device, compared with the most recent devices.
More Related Videos
06:57Use of an Integrated Low-Flow Anesthetic Vaporizer, Ventilator, and Physiological Monitoring System for Rodents
Published on: July 9, 2020
05:43Topical Airway Anesthesia for Awake-endoscopic Intubation Using the Spray-as-you-go Technique with High Oxygen Flow
Published on: January 13, 2017
Related Concept Videos
Administering Oxygen by Nasal Cannula
Nasal Cannulas
A nasal cannula is a lightweight tube split into two prongs placed in the nostrils,...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...
Tracheostomy Decannulation
Description of the Procedure
Decannulation refers to the permanent removal of the tracheostomy tube, signaling the resolution of the condition that initially necessitated the tracheostomy. The process requires a well-coordinated interplay between...
Administering Oxygen by Mask
Administering oxygen by mask is a common nursing intervention that provides supplemental oxygen to patients with respiratory distress or chronic lung conditions. This procedure involves delivering oxygen at a specified rate through a face mask connected to an oxygen source.
Equipment
The equipment necessary for this procedure includes: