Fixation Pins Increase the Accuracy of Implant Surgery in Free-End Models: An In Vitro Study
Xi Chen1, Zhenyu Yang1, Yuan Wang1
1Resident, Stomatological Hospital of Chongqing Medical University, Chongqing Key Laboratory of Oral Diseases and Biomedical Sciences, Chongqing Municipal Key Laboratory of Oral Biomedical Engineering of Higher Education, Chongqing, China.
Purpose:
Implant surgical guides, in combination with implant planning software, have been designed for accurate surgery, especially in partial edentulism. The purpose of this study was to examine the effect of fixation pins of surgical guides on the accuracy of static computer-assisted implant surgeries in a maxillary free-end situation.
Materials And Methods:
This in vitro study was conducted to compare surgical guides using various fixation pin protocols in implant surgery. A patient dental model with missing teeth from maxillary right first premolar to third molar was used as the study model… Implant placement was planned at maxillary right first premolar, right first molar, and right second molar; Straumann full guide templates were designed and fabricated using the coDiagnostiX software. The experiment involved surgical guides with no fixation pins (NF), buccal unilateral fixation pin (BF), palatal unilateral fixation pin (PF), and bilateral fixation pins on the buccal and palatal sides (BPF), based on the position and number of fixation pins. The deviation between the actual and planned positions was used to evaluate implant accuracy. The fixation pin protocols were the primary predictive variables. Angular, 3D platform, and 3D apex deviations were the primary outcome variables. Statistical analysis was performed using the one-way analysis of variance and Tukey's test (α = 0.05).
Results:
NF generated the maximum angular deviation (3.65 ± 1.39°), 3D platform deviation (1.58 ± 0.55 mm), and 3D apex deviation (2.18 ± 0.79 mm), whereas BPF produced the minimum angular deviation (1.88 ± 0.86°), 3D platform deviation (1.09 ± 0.51 mm), and 3D apex deviation (1.53 ± 0.45 mm). A statistically significant difference between NF and BPF in the angular deviation, 3D platform, and apex deviation (P < .0001, P = .009, and P = .002, respectively) was identified. The unilateral fixation pin exerted a significant effect only on the angular accuracy (BF, P = .0018; PF, P = .0001).
Conclusion:
In a maxillary free-end situation, templates with a fixation pin generate better implant accuracy than those without it. A bilateral fixation pin protocol may produce less deviation than those without fixation pins. The implant accuracy does not appear to be affected by the position of the unilateral fixation pin.