Study on drug substance degradation under headspace GC conditions of residual solvent analysis
Hua Zhao1, Doug Phillipson2, Jiangwei Li1
1Neurocrine Biosciences Inc, 12780 El Camino Real, San Diego, CA 92130, USA. lzhao@neurocrine.com.
Abstract:
Isobutyraldehyde (IBA) was detected in drug substance (DS) containing an amino acid group using a headspace-gas chromatography (HS-GC) method. High-performance liquid chromatography-mass-spectrometry (HPLC-MS) data from an HS vial confirmed that IBA was a degradant. The HS-GC method was modified to minimize IBA by keeping the HS oven temperature lower than 80 °C.
More Related Videos
05:29Profiling Volatile Compounds in Blackcurrant Fruit using Headspace Solid-Phase Microextraction Coupled to Gas Chromatography-Mass Spectrometry
Published on: June 9, 2021
07:57Quantitative Detection of Trace Explosive Vapors by Programmed Temperature Desorption Gas Chromatography-Electron Capture Detector
Published on: July 25, 2014
Related Concept Videos
Gas Chromatography: Introduction
In GC, a sample is vaporized and mixed with an inert carrier gas (the mobile phase), which transports it through a...
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall....
Supercritical Fluid Chromatography
SFC utilizes a supercritical fluid mobile phase,...
Sample Preparation for Analysis: Overview
Bulk or large solid samples are typically reduced in size using grinding, crushing, or milling techniques to increase the...
Volatilization
Gas Chromatography: Sample Injection Systems
Two primary injection methods are used...