Related Experiment Video
Updated: Aug 12, 2025

A Microfluidic Flow Chamber Model for Platelet Transfusion and Hemostasis Measures Platelet Deposition and Fibrin Formation in Real-time
Published on: February 14, 2017
Short-Term (6 Weeks) Experience of a Modular Workcell for Hemostasis Testing Including an Intelligent Data Manager at
Hwan Tae Lee1,2, So Young Lee1, Ja Young Seo1
1Department of Laboratory Medicine, Gil Medical Center, Gachon University College of Medicine, Incheon, Republic of Korea.
Objective:
Modular workcells could be a better solution than total laboratory automation (TLA) in hemostasis laboratories. Here, we evaluated the impact of implementing a modular workcell (HemoCell) with an intelligent data management facility (HemoHub).
Methods:
We compared the turnaround times (TATs), numbers of rerun samples, and rerun times pre- and postimplementation of the HemoCell at Gil Medical Center. Prothrombin time (PT), activated partial thromboplastin time (aPTT), D-dimer, and fibrinogen were evaluated.
Results:
The TAT standard deviations (SDs) and maximum TAT values decreased after HemoCell implementation, although the mean TATs for PT, aPTT, and D-dimer were increased. Numbers of rerun samples were increased (18.1/day vs 44.7/day). However, rerun times were reduced, and SDs were decreased during the post-HemoCell period compared with pre-HemoCell. Additionally, technologists needed smaller working space and less labor.
Conclusion:
The modular workcell could improve quality and efficiency by providing more consistent TATs and shorter rerun times in the hemostasis laboratory.