Effects of using in-line bacterial filters with continuous positive airway pressure devices in patients with sleep
Satoshi Hamada1, Jumpei Togawa2, Hironobu Sunadome2
1Department of Advanced Medicine for Respiratory Failure, Graduate School of Medicine, Kyoto University, Kyoto, Japan.
Abstract:
In June 2021, Philips Respironics first advised the use of in-line bacterial filters in response to a recall of continuous positive airway pressure (CPAP) devices. However, the effects of using in-line filters with CPAP devices have not yet been examined in a clinical setting. Here, we retrospectively evaluated 160 patients with sleep-disordered breathing (SDB) who voluntarily used in-line bacterial filters. Patients' perceptions of CPAP pressure after using in-line filters were investigated via a verbal interview. Patients described CPAP pressure as very weak (13, 8.1%), somewhat weak (42, 26.2%), unchanged (103, 64.4%), or somewhat strong (2, 1.3%). Five of the thirteen patients (38.5%) who described CPAP pressure as very weak discontinued the use of in-line filters. The apnea-hypopnea index obtained from the CPAP devices did not change after using in-line filters (n = 129, p = 0.85). Hence, although in-line bacterial filters could change patients' perceptions of CPAP pressure, they did not influence SDB control.
Insights
In-line bacterial filters for CPAP machines did not affect sleep-disordered breathing control. While some patients perceived pressure changes, the filters did not alter apnea-hypopnea index, indicating no impact on SDB management.
Area of Science:
- Respiratory Medicine
- Sleep Medicine
- Medical Device Technology
Background:
- Philips Respironics recommended in-line bacterial filters for continuous positive airway pressure (CPAP) devices in June 2021 due to a recall.
- The clinical impact of using these in-line filters with CPAP devices remained unexamined.
Purpose of the Study:
- To evaluate the effects of in-line bacterial filters on CPAP therapy in patients with sleep-disordered breathing (SDB).
- To assess patient perceptions of CPAP pressure when using in-line filters.
- To determine if in-line filters influence SDB control as measured by the apnea-hypopnea index.
Main Methods:
- Retrospective evaluation of 160 patients with SDB who voluntarily used in-line bacterial filters.
- Patient perceptions of CPAP pressure were collected through verbal interviews.
- Apnea-hypopnea index (AHI) data from CPAP devices before and after filter use were analyzed.
Main Results:
- A significant portion of patients reported perceived changes in CPAP pressure: 8.1% felt it was very weak, 26.2% somewhat weak, 64.4% unchanged, and 1.3% somewhat strong.
- Among those perceiving very weak pressure, 38.5% discontinued filter use.
- No significant change in the apnea-hypopnea index was observed after using in-line filters (n=129, p=0.85).
Conclusions:
- In-line bacterial filters may alter subjective patient perception of CPAP pressure.
- These filters do not appear to impact the objective control of sleep-disordered breathing as indicated by the apnea-hypopnea index.
- The use of in-line filters in CPAP therapy for SDB requires further investigation regarding patient adherence and objective efficacy.
More Related Videos
04:53Author Spotlight: Traditional Chinese Medicine for Sleep Disorders in Acute COPD — A Safe, Cost-Effective Approach
Published on: October 18, 2024
07:54Drug-Induced Sleep Endoscopy DISE with Target Controlled Infusion TCI and Bispectral Analysis in Obstructive Sleep Apnea
Published on: December 6, 2016
Related Concept Videos
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Sleep Apnea
The condition is more prevalent among...
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Mechanical Ventilation I: Indication and Settings
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...