Effects of using in-line bacterial filters with continuous positive airway pressure devices in patients with sleep disordered breathing
Related Experiment Video
Updated: Aug 11, 2025

Effects of Surgical Masks on Cardiopulmonary Function in Healthy Subjects
Published on: February 12, 2021
Satoshi Hamada1, Jumpei Togawa2, Hironobu Sunadome2
1Department of Advanced Medicine for Respiratory Failure, Graduate School of Medicine, Kyoto University, Kyoto, Japan.
Abstract:
In June 2021, Philips Respironics first advised the use of in-line bacterial filters in response to a recall of continuous positive airway pressure (CPAP) devices. However, the effects of using in-line filters with CPAP devices have not yet been examined in a clinical setting. Here, we retrospectively evaluated 160 patients with sleep-disordered breathing (SDB) who voluntarily used in-line bacterial filters. Patients' perceptions of CPAP pressure after using in-line filters were investigated via a verbal interview. Patients described CPAP pressure as very weak (13, 8.1%), somewhat weak (42, 26.2%), unchanged (103, 64.4%), or somewhat strong (2, 1.3%). Five of the thirteen patients (38.5%) who described CPAP pressure as very weak discontinued the use of in-line filters. The apnea-hypopnea index obtained from the CPAP devices did not change after using in-line filters (n = 129, p = 0.85). Hence, although in-line bacterial filters could change patients' perceptions of CPAP pressure, they did not influence SDB control.
Insights
In-line bacterial filters for CPAP machines did not affect sleep-disordered breathing control. While some patients perceived pressure changes, the filters did not alter apnea-hypopnea index, indicating no impact on SDB management.
Area of Science:
- Respiratory Medicine
- Sleep Medicine
- Medical Device Technology
Background:
- Philips Respironics recommended in-line bacterial filters for continuous positive airway pressure (CPAP) devices in June 2021 due to a recall.
- The clinical impact of using these in-line filters with CPAP devices remained unexamined.
Purpose of the Study:
- To evaluate the effects of in-line bacterial filters on CPAP therapy in patients with sleep-disordered breathing (SDB).
- To assess patient perceptions of CPAP pressure when using in-line filters.
- To determine if in-line filters influence SDB control as measured by the apnea-hypopnea index.
Main Methods:
- Retrospective evaluation of 160 patients with SDB who voluntarily used in-line bacterial filters.
- Patient perceptions of CPAP pressure were collected through verbal interviews.
- Apnea-hypopnea index (AHI) data from CPAP devices before and after filter use were analyzed.
Main Results:
- A significant portion of patients reported perceived changes in CPAP pressure: 8.1% felt it was very weak, 26.2% somewhat weak, 64.4% unchanged, and 1.3% somewhat strong.
- Among those perceiving very weak pressure, 38.5% discontinued filter use.
- No significant change in the apnea-hypopnea index was observed after using in-line filters (n=129, p=0.85).
Conclusions:
- In-line bacterial filters may alter subjective patient perception of CPAP pressure.
- These filters do not appear to impact the objective control of sleep-disordered breathing as indicated by the apnea-hypopnea index.
- The use of in-line filters in CPAP therapy for SDB requires further investigation regarding patient adherence and objective efficacy.
More Related Videos
04:53Author Spotlight: Traditional Chinese Medicine for Sleep Disorders in Acute COPD — A Safe, Cost-Effective Approach
Published on: October 18, 2024
07:54Drug-Induced Sleep Endoscopy DISE with Target Controlled Infusion TCI and Bispectral Analysis in Obstructive Sleep Apnea
Published on: December 6, 2016
Related Concept Videos
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Sleep Apnea
The condition is more prevalent among...
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Mechanical Ventilation I: Indication and Settings
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...