Bodyweight-adjusted Contrast Media With Shortened Injection Duration for Step-and-Shoot Coronary Computed Tomography
Liang Jin1, Kun Wang1, Xiaodong Wang2
1Radiology Department, Huadong Hospital, Affiliated with Fudan University.
Purpose:
Shortened injection durations are not recommended in step-and-shoot coronary computed tomography angiography (CCTA). We aimed to evaluate the image quality of CCTA performed using bodyweight-adjusted iodinated contrast media (ICM) with different injection durations to generate an optimized ICM administration protocol to acquire convincible image quality in step-and-shoot CCTA.
Materials And Methods:
A total of 200 consecutive patients with suspected coronary artery disease (CAD) were enrolled in group A (N=50, 350 mgI/mL, bodyweight×0.8 mL/kg with a 13-s injection duration), group B (N=50, 350 mgI/mL, bodyweight×0.9 mL/kg with a 13-s injection duration), group C (N=50, 350 mgI/mL, bodyweight×0.8 mL/kg with a 12-s injection duration), and group D (N=50, 320 mgI/mL, bodyweight×0.8 mL/kg with a 13-s injection duration). Patient characteristics, ICM administration protocols, quantitative computed tomography (CT) value measurements, and qualitative image scores were analyzed and compared among the groups.
Results:
Groups A and D achieved the lowest ICM volume, saline volume, injection flow rate, and total iodine and iodine injection rates among the groups. All the CT values of the coronary arteries in all groups were >300 HU. All the observers' average scores exceeded three points. In group A, the CT values showed significant positive correlation with the iodine injection rate ( r =0.226, P <0.001), whereas the signal-to-noise ratio ( r =-0.004, P =0.927) and contrast-to-noise ratio ( r =-0.006, P =0.893) values were not.
Conclusions:
Bodyweight×0.8 mL/kg with a 13-second injection duration is a comprehensive option for step-and-shoot CCTA with improved image quality, and a 350 mgI/mL iodine concentration is preferred.
More Related Videos
08:13In Vivo Quantitative Assessment of Myocardial Structure, Function, Perfusion and Viability Using Cardiac Micro-computed Tomography
Published on: February 16, 2016
09:32Time-Resolved, Dynamic Computed Tomography Angiography for Characterization of Aortic Endoleaks and Treatment Guidance via 2D-3D Fusion-Imaging
Published on: December 9, 2021
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Imaging Studies III: Computed Tomography
Radiological Investigation I: X-ray and CT
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System I:Echocardiography
Indications: Echocardiography is utilized to diagnose heart failure, valve disorders, and myocardial infarction. It also assesses cardiac structures' size, shape, and motion,...
Imaging Studies for Cardiovascular System IV: CMRI