Introduction:
Carcasses from animal experiments with RG-3 pathogens should be decontaminated onsite in Austria.
Objective:
The aim of this study was to find out if the use of pass-through autoclaves for the decontamination of animal carcasses (up to 40 kg of weight) could serve as a routine method for smaller laboratories, as the installation of special carcass decontamination plants may be cost prohibitive.
Methods:
Biological indicators (BIs) were implanted into the carcasses of animals of different sizes and species with a novel method using stainless steel pipes. The bodies were placed in autoclavable plastic bags and equipped with thermal probes by insertion through the rectum. Subsequently a factory default autoclave cycle for liquids was performed, which holds a core temperature of 121°C for 20 min.
Results:
The weight of the carcasses ranged from 1 to 42 kg, the duration of the individual cycles reached from 2.2 to 17.23 h. Decontamination was successful every single time as shown by the BIs. The application through the natural orifices with the help of the application tools seems to offer a reliable alternative for implanting the BIs into the carcasses without creating new openings. Insulation properties did not pose substantial challenges to the process. Limitations on the packaging procedure were identified in carcasses larger than 30 kg.
Conclusion:
Based on the results of this study, using pass-through autoclaves represents an option as a routine method for the decontamination of animal carcasses up to at least 40 kg.
Related Concept Videos
Methods of Sterilization I: Physical Methods
Steam sterilization uses non-toxic, low-cost moist heat in the form of saturated steam under pressure, which is fast, microbicidal, and sporicidal, and quickly warms and penetrates fabrics. Autoclaves, or steam sterilizers, expose each item to direct steam contact for a predetermined time at the necessary...
Methods of Sterilization II: Chemical Methods
Using chemical sterilization rather than heat to clean out equipment is recommended. It eradicates and removes all bacteria,...
Physical Methods for Controlling Microbial Growth: Temperature
Cleaning, Sterilization, and Disinfection
Cleaning
The cleaning process usually involves using water with detergents or enzymatic cleaner and removing foreign material from objects and surfaces, including organic material such as body fluids or inorganic material like soil. Cleaning is performed before high-level disinfection and sterilization because foreign materials on the cover of the devices interfere with process...