Assessing Cortical Cerebral Microinfarcts on High Resolution MR Images
Published on: November 20, 2015
Prospective Assessment of Cerebral Microbleeds with Low-Field Magnetic Resonance Imaging (0.55 Tesla MRI)
Thilo Rusche1, Hanns-Christian Breit1, Michael Bach1
1Department of Radiology, Clinic of Radiology & Nuclear Medicine, University Hospital of Basel, University of Basel, 4001 Basel, Switzerland.
Purpose:
Accurate detection of cerebral microbleeds (CMBs) on susceptibility-weighted (SWI) magnetic resonance imaging (MRI) is crucial for the characterization of many neurological diseases. Low-field MRI offers greater access at lower costs and lower infrastructural requirements, but also reduced susceptibility artifacts. We therefore evaluated the diagnostic performance for the detection of CMBs of a whole-body low-field MRI in a prospective cohort of suspected stroke patients compared to an established 1.5 T MRI.
Methods:
A prospective scanner comparison was performed including 27 patients, of whom 3 patients were excluded because the time interval was >1 h between acquisition of the 1.5 T and 0.55 T MRI. All SWI sequences were assessed for the presence, number, and localization of CMBs by two neuroradiologists and additionally underwent a Likert rating with respect to image impression, resolution, noise, contrast, and diagnostic quality.
Results:
A total of 24 patients with a mean age of 74 years were included (11 female). Both readers detected the same number and localization of microbleeds in all 24 datasets (sensitivity and specificity 100%; interreader reliability ϰ = 1), with CMBs only being observed in 12 patients. Likert ratings of the sequences at both field strengths regarding overall image quality and diagnostic quality did not reveal significant differences between the 0.55 T and 1.5 T sequences (p = 0.942; p = 0.672). For resolution and contrast, the 0.55 T sequences were even significantly superior (p < 0.0001; p < 0.0003), whereas the 1.5 T sequences were significantly superior (p < 0.0001) regarding noise.
Conclusion:
Low-field MRI at 0.55 T may have similar accuracy as 1.5 T scanners for the detection of microbleeds and thus may have great potential as a resource-efficient alternative in the near future.
Insights
Low-field MRI (0.55 T) shows comparable accuracy to 1.5 T MRI for detecting cerebral microbleeds. This offers a cost-effective alternative for neurological disease characterization.
Area of Science:
- Radiology
- Neurology
- Medical Imaging
Background:
- Cerebral microbleeds (CMBs) detection is vital for diagnosing neurological diseases.
- Susceptibility-weighted imaging (SWI) on MRI is key for CMB detection.
- Low-field MRI offers accessibility and cost benefits but faces challenges with artifacts.
Purpose of the Study:
- To evaluate the diagnostic performance of a whole-body low-field (0.55 T) MRI for CMB detection.
- To compare the low-field MRI's accuracy against a standard 1.5 T MRI in suspected stroke patients.
Main Methods:
- Prospective comparison of 0.55 T and 1.5 T MRI SWI sequences in 24 patients.
- Two neuroradiologists assessed CMB presence, number, and location.
- Image quality, resolution, noise, and contrast were rated using a Likert scale.
Main Results:
- Identical detection of CMBs (number and location) by both readers across all datasets (100% sensitivity and specificity).
- No significant differences in overall image and diagnostic quality between 0.55 T and 1.5 T.
- 0.55 T showed superior resolution and contrast, while 1.5 T had less noise.
Conclusions:
- Low-field 0.55 T MRI demonstrates similar accuracy to 1.5 T MRI for detecting cerebral microbleeds.
- 0.55 T MRI presents a promising, resource-efficient alternative for microbleed detection.