Comparison of image quality and radiation dose of different scanning methods used for computed tomography of the unilateral shoulder
Related Experiment Video
Updated: Aug 10, 2025

Measuring 3D In-vivo Shoulder Kinematics using Biplanar Videoradiography
Published on: March 12, 2021
Dongdong Jiang1, Yanan Wang2, Peng Zhang1
1Department of Radiology, Wuhan Fourth Hospital, Wuhan, Hubei, PR China.
Background:
The effect of different computed tomography (CT) scanning methods of the shoulder on image quality is uncertain.
Purpose:
To compare the effect of different methods of CT scanning of the right shoulder on image quality and radiation dose.
Material And Methods:
A total of 30 adults were divided into five groups. Group A received scans centered on the body's long axis, a scout direction of 0° + 90°, and automatic tube current modulation (ATCM). The other four groups (B, C, D, E) received isocenter scans centered on the shoulder with different scout directions (B and C: 0° + 90°, D: 0°, E: 0° + 270°) and tube currents (B: 420 mA; C, D, E: ATCM). The volume CT dose index (CTDIvol), dose-length product (DLP), image objective noise, signal-to-noise ratio (SNR), and contrast-to-noise ratio (CNR) were compared. Three subjective measures were also compared (noise, stripe artifacts, diagnostic confidence).
Results:
The five groups differed significantly in all subjective and objective indexes. The CTDIvol and DLP decreased in the order of groups C, A, B, E, and D; the differences between groups A and B were not significant (P > 0.05). Groups B, C, and E had better SNR and CNR than groups A and D (P < 0.01). Subjective evaluations indicated group D was worse than groups B, C, and E (P < 0.05).
Conclusion:
In the ATCM system that uses the last scout view, CT of the shoulder should use isocenter scanning with the lateral scout view when the tube is away from the long axis of the body as the last execution direction.
More Related Videos
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Ultrasonography
During an ultrasonography procedure, a handheld device called...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Radiological Investigation I: X-ray and CT
Imaging Studies for Cardiovascular System V: CT