Validation of Linear Range HER2/Estrogen Receptor/Progesterone Receptor IHControls for Daily Quality Assurance
Seshi R Sompuram1, Kodela Vani1, Lori Ryan2
1Boston Cell Standards, Boston MA, USA.
American Journal of Clinical Pathology
|February 13, 2023
Summary
The new FDA-cleared IHControls demonstrate high concordance with traditional tissue controls for HER2, ER, and PR testing. These controls accurately detect immunohistochemistry (IHC) assay issues, offering reliable performance in clinical settings.
Area of Science:
- Biomedical diagnostics
- Immunohistochemistry
- Quality control
Background:
- Immunohistochemistry (IHC) is crucial for cancer diagnostics, relying on accurate controls.
- Traditional tissue controls can have variability and limitations.
- A new FDA-cleared control, IHControls, utilizes peptide epitopes on microspheres.
Purpose of the Study:
- To evaluate the performance of IHControls for HER2, estrogen receptor (ER), and progesterone receptor (PR).
- To compare IHControls against conventional tissue controls in clinical IHC laboratories.
- To assess the sensitivity, reproducibility, and specificity of IHControls.
Main Methods:
- Side-by-side comparison of IHControls and tissue/cell line controls across 5 clinical labs.
- Simulated IHC assay failure experiments using primary antibody dilution.
- Evaluation of lot-to-lot manufacturing reproducibility and specificity against irrelevant stains.
Main Results:
- Achieved 99.6% concordance between IHControls and tissue controls across 766 evaluations.
- Both control types demonstrated similar sensitivity in detecting IHC assay errors.
- IHControls showed <10% variability in manufacturing reproducibility and no cross-reactivity.
Conclusions:
- IHControls are the first FDA-cleared IHC controls.
- They offer sensitive and accurate detection of IHC assay problems.
- Performance is comparable to traditional tissue controls.