Clinical Retinal Image Quality of a Non-diffractive Wavefront-Shaping Extended Depth of Focus (Vivity) Intraocular
Purpose:
To evaluate clinical retinal optical image quality following implantation of an extended depth of focus intraocular lens (EDOF IOL) (Vivity; Alcon Laboratories, Inc), and to compare it with a monofocal and a trifocal IOL.
Methods:
This prospective, comparative, case-control study included 88 eyes implanted with: (1) 19 monofocal IOLs (AcrySof SA60AT; Alcon Laboratories, Inc); (2) 38 EDOF IOLs (AcrySof IQ Vivity); and (3) 31 trifocal IOLs (AT LISA tri 839MP; Carl Zeiss Meditec AG). Total root mean square, ocular lower (LOA) and higher (HOA) order aberrations, point spread function (PSF) Strehl ratio (PSF with LOA), and PSF Strehl ratio excluding LOA (PSF without LOA) were analyzed using a Pyramidal WaveFront-based sensor aberrometer Osiris (Costruzione Strumenti Oftalmici) at two different pupil sizes (3 and 4 mm).
Results:
The trifocal IOL showed the highest PSF without LOA at both pupil sizes (0.52 ± 0.12 and 0.31 ± 0.07, respectively), followed by the AcrySof SA60AT (0.39 ± 0.10 and 0.27 ± 0.07) and AcrySof IQ Vivity (0.34 ± 0.11 and 0.24 ± 0.09) (P < .001). The AcrySof IQ Vivity and monofocal IOLs were comparable (P > .05). Despite the comparable postoperative low spherical equivalent among the IOL groups, the AT LISA tri 839MP retinal image quality (PSF with LOA) was the most severely affected by such residual refractive errors (dropped to 0.26 ± 0.06 at 3 mm; P < .001) compared to the monofocal AcrySof SA60AT (0.24 ± 0.07 at 3 mm) and EDOF Acrysof IQ Vivity (0.23 ± 0.06 at 3 mm) groups. The PSF with LOA was comparable (P > .05) among the three groups at both the 3-and 4-mm pupil size.
Conclusions:
Although trifocal IOLs provided significantly better retinal image quality if influence of LOA is excluded, they also demonstrated to be the most sensitive to residual refractive errors. Both the EDOF Acrysof IQ Vivity and mono-focal AcrySof SA60AT IOLs showed a comparable retinal image quality, and they are also comparable with trifocal IOLs when considering the clinically real PSF (PSF with LOA). [.
Related Concept Videos
Focusing of Light in the Eye
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...
Angle Closure Glaucoma: Treatment
Open Angle Glaucoma: Treatment
Drugs such as carbonic anhydrase inhibitors, α2- and...