Validation of the A&D UM-212BLE monitor according to ISO 81060-2, 2018: a device with clinically important
1Department of Pediatrics, University of Tennessee Health Science Center, Memphis, Tennessee, USA.
Objective:
The objective of this report was to describe the validation of the A&D UM-212BLE automated oscillometric sphygmomanometer to the ISO 81060-2, 2018 protocol. The device is specifically designed for enhanced office and out-of-office programmability.
Methods:
A combined pediatric ( n = 35) and adult ( n = 50) population was studied at Clinmark LLC in Louisville, Colorado, USA. Same-arm sequential testing was performed following the ISO 81060-2, 2018 requirements. Five cuffs were tested with a total arm circumference range from 12 to 50 cm. Reference readings were done by two blinded observers performing simultaneous auscultation.
Results:
For validation of Criterion 1 the mean ± SD (mmHg) of the device minus the reference differences were 3.94 ± 6.89 for SBP and 2.09 ± 6.68 for DBP. Both passed the Standard limits for Criterion 2; the systolic(S)SD achieved was 5.56 (5.70 permitted) and the diastolic(D)SD was 6.01 (6.62 permitted). All other Standard requirements were met.
Conclusions:
The UM-212BLE passed all requirements. The features that make this device clinically superior include settings for automated office BP, variable pressure inflation, dual measurement modes (oscillometry, auscultation), the wide range of cuffs tested, automated irregular heartbeat detection, and full validation in a pediatric population. The inclusion of all of these features makes the UM-212BLE a highly attractive device for both office and out-of-office BP estimation.
More Related Videos
Related Concept Videos
Data Validation
Key parameters for method validation include:
Assessing Body Temperature - Oral
Step 1:
Start by practicing proper hand hygiene to prevent the spread of microorganisms.
Step 2:
Take the thermometer out of the charging unit, switch it on, and wait for the ready sign.
Step 3:
Gently slide the probe cover until a click is heard. This simple action prevents cross-contamination and ensures the correct placement of the probe cover.
Step 4:
Instruct the patient to open their mouth and place...
Uncertainty in Measurement: Reading Instruments