A Single-Blind Study Testing the Preparation Accuracy of Bedside Precleaning Solutions Used for Flexible Endoscopes
Vera Meeusen1,2, Traci McLean1,2
1Vera Meeusen, PhD, MA, RN, FACPAN, AFACHSM, is A/Professor; Clinical Nurse Consultant, Endoscopy Unit, Princess Alexandra Hospital, Woolloongabba, Queensland, Australia; and A/Professor in Medicine, The University of Queensland, St Lucia, Queensland, Australia.
Abstract:
Endoscopy-related pathogen transmission may occur if microorganisms are spread from patient to patient by contaminated equipment. Effective and safe endoscope reprocessing includes bedside precleaning, mechanical cleaning, high-level disinfection, storage, and drying. The aim of this research study was to observe and report on the variation in practice regarding the use of detergent for bedside precleaning of flexible gastrointestinal endoscopes. Endoscopy nurses working in the endoscopy unit at the Princess Alexandra Hospital, Australia, prepared four samples of detergent solution as per normal routine. Twenty-nine nurses participated providing in total 116 samples. There was a significant variation in detergent concentration. The detergent concentration variated between 2.00 and 288.20 ml/L ( M = 34.55, SD = 39.21). Two samples revealed lower concentrations than required. More than 25% of the samples contained at least a 10 times higher concentration than required (>40 ml/L). Current practice of bedside precleaning of gastrointestinal endoscopes was not deemed safe or cost-effective as it did not guarantee an adequate concentration of detergent. More precise methods to establish the required concentration of the bedside precleaning solution were introduced to improve practice.
More Related Videos
06:28E-Patient Counseling Trial E-PACO: Computer Based Education versus Nurse Counseling for Patients to Prepare for Colonoscopy
Published on: August 1, 2019
07:00Failure of Cleaning Verification in Pharmaceutical Industry Due to Uncleanliness of Stainless Steel Surface
Published on: August 11, 2017