Real-Time Cardiac Mapping with a Noninvasive Imageless Electrocardiographic Imaging System
Published on: April 11, 2025
The effectiveness and feasibility of using multi-lead ECG monitoring combined with a programmed intracavitary ECG to
Min Wang1,2, Yaxun Sun1,2, Yu Shan1,2
1Department of Cardiology, Key Laboratory of Cardiovascular Intervention and Regenerative Medicine of Zhejiang Province, Hangzhou, China.
Background:
Left bundle branch area pacing (LBBaP) as an alternative method for delivering physiological pacing, is difficult for many primary hospitals that lack the electrophysiological multichannel recorder to carry out. We hope to find a simple and feasible method that combines the multi-lead surface electrocardiogram (ECG) monitoring and the intracavity ECG of the pacing programmer to achieve LBBaP.
Methods:
A total of 50 patients with bradycardia indications who attempted permanent pacemaker implantation were included in this study. We referred to multi-lead surface ECG monitoring and pacing system analyzer (PSA), combined with the nine-zone pacing method of the LBBaP, to complete LBBaP. We assessed multiple parameters to verify whether the LBBaP was successfully achieved and used univariable analysis of variance for repeated measures to judge the feasibility and effectiveness of LBBaP without the electrophysiological multichannel recorder.
Results:
LBBaP was successfully archived without the electrophysiological multichannel recorder in 44 of 50 patients (88%). In the study, paced QRS duration and the stimulus to peak left ventricular activation time (Sti-LVAT) were 117.04 ± 10.34 ms and 71.10 ± 7.91 ms and had no significant changes in the 3-month follow-up. The unipolar pacing threshold and R-wave amplitudes were 0.85 ± 0.32 V and 10.36 ± 5.24 mV at baseline respectively, which also showed stability during the 1-month and 3-month follow-up. During the 3-month follow-up, no lead-related complication was recorded.
Conclusion:
It is effective and feasible to achieve LBBaP combining the multi-lead ECG monitoring and the intracavitary ECG of PSA without the electrophysiological multichannel recorder, which could be an alternative to perform LBBaP.
More Related Videos
12:45Benefits of Cardiac Resynchronization Therapy in an Asynchronous Heart Failure Model Induced by Left Bundle Branch Ablation and Rapid Pacing
Published on: December 11, 2017
10:05Testing the Efficacy of Pharmacological Agents in a Pericardial Target Delivery Model in the Swine
Published on: July 7, 2016
Related Concept Videos
Electrocardiogram Fundamentals
An electrocardiogram (ECG) is a diagnostic tool for identifying cardiac conditions such as arrhythmias, conduction abnormalities, and myocardial ischemia.
Definition
An electrocardiogram (ECG) visualizes the heart's electrical activity by tracing the electrical movement associated with each heartbeat on a graph or monitor. As the heart beats, an electrical wave passes through it, correlating with the cardiac cycle events.
Parts of an ECG
An ECG utilizes electrodes on the skin...
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...