Ultrasound Based Assessment of Coronary Artery Flow and Coronary Flow Reserve Using the Pressure Overload Model in Mice
Published on: April 13, 2015
Computing intracoronary blood flow rate under incomplete boundary conditions: Combing coronary anatomy and fractional
Weihao Li1, Kun Lian2, Yan Chen3
1Department of Information Engineering, School of Electronics and Information, Northwestern Polytechnical University, Xi'an, Shaanxi, P.R.China.
Insights
A new computational fluid dynamic method accurately measures intracoronary blood flow rate by including side branches. This improves diagnosis of ischemic heart disease (IHD) compared to methods that ignore these crucial pathways.
Area of Science:
- Biomedical Engineering
- Cardiovascular Physiology
- Medical Imaging Analysis
Background:
- Accurate measurement of intracoronary blood flow is critical for diagnosing ischemic heart disease (IHD).
- Current computational fluid dynamic (CFD) methods using fractional flow reserve (FFR) have limitations due to incomplete boundary conditions, often ignoring side branches.
- This omission can significantly impact the accuracy of mean flow rate calculations.
Purpose of the Study:
- To propose a novel CFD-based method for calculating mean intracoronary blood flow rate.
- To enhance accuracy by incorporating side branches into the model, addressing limitations of existing methods under incomplete pressure boundary conditions.
Main Methods:
- Employed a pressure-flow curve-based flow resistance model for epicardial arteries.
- Utilized steady flow simulations to extract parameters for the flow resistance model.
- Developed a method to calculate mean intracoronary blood flow rate in multiple branches under incomplete pressure boundary conditions by implicitly constraining flow splitting.
Main Results:
- The proposed method demonstrated significantly lower mean percentage errors: +2.05%±0.04% for idealized models and +2.24%±0.01% for reconstructed models.
- In contrast, methods ignoring side branches showed much higher errors (+38.48%±10.45% and +30.54%±6.12%, respectively).
- Even with inconsistent Murray's Law coefficients, percentage errors remained below approximately 3.00%.
Conclusions:
- The developed CFD method offers an accurate and accessible approach for measuring mean intracoronary blood flow rate.
- This advancement is expected to facilitate more precise diagnosis of ischemic heart disease (IHD).
Background:
Accurate measurement of intracoronary blood flow rate is of great significance for the diagnosis of ischemic heart disease (IHD). Computational fluid dynamic (CFD) method, combining coronary angiography images and fractional flow reserve (FFR), provides a new way to calculate the mean flow rate. However, due to the incomplete boundary conditions obtained by FFR, side branches were ignored which was likely to have a significant impact on the accuracy. In this paper, a novel CFD based method for calculating the mean intracoronary flow rate under incomplete pressure boundary conditions was proposed, in order to improve the accuracy by including the side branches.
Methods:
A pressure-flow curve based flow resistance model was employed to model resistance of the epicardial arteries. A series of steady flow simulations were performed to extract the parameters of the flow resistance model, which implicitly specified constraints for splitting flow between branches and thus enabled the mean intracoronary blood flow rate to be calculated in two or more branches under incomplete pressure boundary conditions. Simulation experiments were designed to validate the proposed method in both idealized and reconstructed 3D models of coronary branches, and the impact of the assumed coefficient of the Murray's Law for splitting flow between branches was also investigated.
Results:
The mean percentage error of the proposed method was +2.05%±0.04% for idealized models and +2.24%±0.01% for reconstructed models, and it was much lower than that of the method ignoring side branches (+38.48%±10.45% for idealized models and +30.54%±6.12% for reconstructed models). When the assumed coefficient of the Murray's Law was inconsistent with the real blood flow condition, the percentage errors still maintained less than about 3.00%.
Conclusions:
The proposed method provided an easy and accurate way to measure the mean intracoronary flow rate and would facilitate the accurate diagnosis of IHD.