Modified design of x-ray protective clothing to enhance radiation protection for interventional radiologists
Tae Won Choi1, Jin Wook Chung2,3,4, Yohan Kwon1
1Department of Radiology, Ajou University School of Medicine, Suwon, Republic of Korea.
Background:
In interventional radiology procedures, the operator typically stands on the right side of the patient's right thigh to manipulate devices through the femoral sheath. Because the standard x-ray protective clothing is designed as sleeveless and scatter radiations from the patient are mainly incident from the left-anterior direction to the operator, the arm hole of the clothing may be a significant unprotected area, contributing to an increase in the operator's organ doses and effective dose.
Purpose:
This study aimed to compare the organ doses and effective dose received by the interventional radiologist when wearing the standard x-ray protective clothing and when wearing the modified clothing with an additional shoulder guard.
Methods:
The experimental setup aimed to simulate actual clinical practice in interventional radiology. The patient phantom was located at the beam center to generate scatter radiation. An adult female anthropomorphic phantom loaded with 126 nanoDots (Landauer Inc., Glenwood, IL) was used to measure organ and effective doses to the operator. The standard wrap-around type x-ray protective clothing offered 0.25-mm lead-equivalent protection, and the frontal overlap area offered 0.50-mm lead-equivalent protection. The shoulder guard was custom-made with a material providing x-ray protection equivalent to lead of 0.50 mm thickness. The organ and effective doses were compared between the operator wearing the standard protective clothing and the one wearing the modified clothing with a shoulder guard.
Results:
After adding the shoulder guard, doses to the lungs, bone marrow, and esophagus decreased by 81.9%, 58.6%, and 58.7%, respectively, and the effective dose to the operator decreased by 47.7%.
Conclusions:
Widespread use of modified x-ray protective clothing with shoulder guards can significantly decrease the overall occupational radiation risk in interventional radiology.
More Related Videos
12:57X-Ray Visualization of Intraductal Ethanol-Based Ablative Treatment for Prevention of Breast Cancer in Rat Models
Published on: December 9, 2022
13:48Reduction of Radiation Exposure during Endovascular Treatment of Peripheral Arterial Disease Combining Fiber Optic RealShape Technology and Intravascular Ultrasound
Published on: April 21, 2023
Related Concept Videos
Radiological Investigation I: X-ray and CT
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
X-ray Imaging
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies for Cardiovascular System V: CT
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...