Validation of e-Celsius gastrointestinal telemetry system as measure of core temperature
O C Koumar1, R Beaufils1, C Chesneau2
1Normandie Univ, UNICAEN, INSERM, COMETE, GIP CYCERON, 14000, Caen, France.
Abstract:
The main objective of this study was to validate gastrointestinal measurement with the e-Celsius® system composed of an ingestible electronic capsule and a monitor. Twenty-three healthy volunteers aged 18-59 years stayed at the hospital for 24 h under fasting conditions. They were only allowed for quiet activity and were asked to keep their sleeping habits. Subjects ingested a Jonah capsule and an e-Celsius® capsule, and a rectal probe and an esophageal probe were inserted. Mean temperature measured by the e-Celsius® device was lower than that measured by Vitalsense® (-0.12 ± 0.22°C; p < 0.001) and the rectal probe (-0.11 ± 0.03°C; p = 0.003) and higher than that measured by the esophageal probe (0.17 ± 0.05; p = 0.006). Mean difference (bias) and 95% confidence intervals between temperature of e-Celsius capsule, Vitalsense Jonah capsule, esophageal probe, and rectal probe were computed using Bland and Altman procedure. The magnitude of the measurement bias is significantly greater when comparing the e-Celsius® and the Vitalsense® device pair with any other device pairs containing the esophageal probe. Amplitude of confidence interval between the e-Celsius® system and the Vitalsense® system was 0.67°C. This amplitude was significantly lower than those of the esophageal probe-e-Celsius® pairing (0.83°C; p = 0.027), of the esophageal probe-Vitalsense (0.78°C; p = 0.046) and of the esophageal probe-rectal probe (0.83°C; p = 0.002). The statistical analysis did not reveal any effect of time on the amplitude of bias, whatever the device concerned. When comparing missing data rate of the e-Celsius® system (0.23 ± 0.15%) and the Vitalsense® devices (0.70 ± 0.11%) during the whole experiment, no differences was observed (p = 0.09). The e-Celsius® system could be used when a continuous following of internal temperature is needed.
More Related Videos
05:43Author Spotlight: Improving Lesion Contiguity in Pulmonary Vein Isolation via Proactive Esophageal Cooling
Published on: April 19, 2024
04:58Reduced Procedure Time and Variability with Active Esophageal Cooling During Radiofrequency Ablation for Atrial Fibrillation
Published on: August 25, 2022
Related Concept Videos
Equipments Used to Measure Body Temperature
Glass-bulb Thermometer:
Glass-bulb thermometers are hollow glass tubes with a bulb tip containing liquid such as ethanol or mercury. Historically, glass bulb mercury thermometers were the standard device to measure body temperature. Today, mercury thermometers are prohibited in many countries due to the hazardous effects of mercury and the risk of exposure if the glass bulb breaks. In general,...
Assessing Body Temperature - Rectal
Follow these steps for rectal temperature assessment:
Step 1: Perform hand hygiene and don clean gloves to prevent cross-infection.
Step 2: Position the patient in a side-lying position to better visualize the rectal...
Assessing Body Temperature - Oral
Step 1:
Start by practicing proper hand hygiene to prevent the spread of microorganisms.
Step 2:
Take the thermometer out of the charging unit, switch it on, and wait for the ready sign.
Step 3:
Gently slide the probe cover until a click is heard. This simple action prevents cross-contamination and ensures the correct placement of the probe cover.
Step 4:
Instruct the patient to open their mouth and place...
Assessing Body Temperature - Axilla
Step 1: Perform hand hygiene and put on clean gloves to maintain infection control and prevent cross-contamination.
Step 2: Prepare the patient by explaining the procedure to ensure understanding and cooperation. Ensure privacy, expose the axilla, and inform the patient that minimal movement is crucial for an accurate reading.
Step 3: Adjust the patient’s clothing to expose only the axilla. It minimizes...
Assessing Body Temperature - Tympanic membrane
Step 1: Begin by practicing good hand hygiene to prevent the transmission of microorganisms.
Step 2: Turn on the thermometer and wait until the ready sign appears on the screen to ensure accurate measurement.
Step 3: Slide the probe cover in place to prevent cross-contamination.
Step 4: Instruct the patient to tilt their head to the side for comfort and check for cerumen...
Assessing Body Temperature - Temporal Artery
Step 1: Perform hand hygiene and don a fresh pair of gloves to prevent cross-infection and ensure patient safety.
Step 2: Explain the procedure to the patient to establish trust. Clear communication establishes trust with the patient, ensures they understand what to expect, promotes cooperation, and enhances comfort during the procedure.
Step 3: Assess the patient's...