1.5-T MR relaxometry in quantifying splenic and pancreatic iron: retrospective comparison of a commercial 3D-Dixon
Michaela Plaikner1, Lukas Lanser2, Christian Kremser3
1Department of Radiology, Medical University of Innsbruck, Anichstraße 35, 6020, Innsbruck, Austria.
Objectives:
To compare the quantitative measurement of splenic and pancreatic iron content using a commercial 3D-Dixon sequence (qDixon) versus an established fat-saturated R2* relaxometry method (ME-GRE).
Methods:
We analyzed splenic and pancreatic iron levels in 143 MR examinations (1.5 T) using the qDixon and a ME-GRE sequence (108 patients: 65 males, 43 females, mean age 61.31 years). Splenic and pancreatic R2* values were compared between both methods using Bland-Altman plots, concordance correlation coefficients (CCC), and linear regression analyses. Iron overload (R2* > 50 1/s) was defined for both organs and compared using contingency tables, overall agreement, and Gwet's AC1 coefficient.
Results:
Of all analyzable examinations, the median splenic R2* using the qDixon sequence was 25.75 1/s (range: 5.6-433) and for the ME-GRE sequence 35.35 1/s (range: 10.9-400.8) respectively. Concerning the pancreas, a median R2* of 29.93 1/s (range: 14-111.45) for the qDixon and 31.25 1/s (range: 14-97) for the ME-GRE sequence was found. Bland-Altman analysis showed a mean R2* difference of 2.12 1/s with a CCC of 0.934 for the spleen and of 0.29 1/s with a CCC of 0.714 for the pancreas. Linear regression for the spleen/pancreas resulted in a correlation coefficient of 0.94 (p < 0.001)/0.725 (p < 0.001). Concerning iron overload, the proportion of overall agreement between the two methods was 91.43% for the spleen and 93.18% for the pancreas.
Conclusions:
Our data show good concordance between R2* values obtained with a commercial qDixon sequence and a validated ME-GRE relaxometry method. The 3D-qDixon sequence, originally intended for liver assessment, seems to be a reliable tool for non-invasive evaluation of iron content also in the spleen and the pancreas.
Key Points:
• A 3D chemical shift imaging sequence and 2D multi-gradient echo sequence show good conformity quantifying splenic and pancreatic R2* values. • The 3D chemical shift imaging sequence allows a reliable analysis also of splenic and pancreatic iron status. • In addition to the liver, the analysis of the spleen and pancreas is often helpful for further differential diagnostic clarification and patient guidance regarding the iron status.