Reducing both radiation and contrast doses for overweight patients in coronary CT angiography with 80-kVp and deep
Wanjiang Li1, Haiyan Lu2, Yuting Wen1
1Department of Radiology, West China Hospital of Sichuan University, Chengdu, Sichuan, China.
Purpose:
To investigate the use of an 80-kVp tube voltage combined with a deep learning image reconstruction (DLIR) algorithm in coronary CT angiography (CCTA) for overweight patients to reduce radiation and contrast doses in comparison with the 120-kVp protocol and adaptive statistical iterative reconstruction (ASIR-V).
Methods:
One hundred consecutive CCTA patients were prospectively enrolled and randomly divided into a low-dose group (n = 50) with 80-kVp, smart mA for noise index (NI) of 36 HU, contrast dose rate of 18 mgI/kg/s and DLIR and 60 % ASIR-V and a standard-dose group (n = 50) with 120-kVp, smart mA for NI of 25 HU, contrast dose rate of 32 mgI/kg/s and 60 % ASIR-V. The radiation and contrast dose, subjective image quality score, attenuation values, noise, signal-to-noise ratio (SNR), and contrast-to-noise ratio (CNR) were compared.
Results:
The low-dose group achieved a significant reduction in the effective radiation dose (1.01 ± 0.45 mSv vs 1.85 ± 0.40 mSv, P < 0.001) and contrast dose (33.69 ± 3.87 mL vs 59.11 ± 5.60 mL, P < 0.001) compared to the standard-dose group. The low-dose group with DLIR presented similar enhancement but lower noise, higher SNR and CNR and higher subjective quality scores than the standard-dose group. Moreover, the same patient comparison in the low-dose group between different reconstructions showed that DLIR images had slightly and consistently higher CT values in small vessels, indicating better defined vessels, much lower image noise, higher SNR and CNR and higher subjective quality scores than ASIR-V images (all P < 0.001).
Conclusions:
The application of 80-kVp and DLIR allows for significant radiation and dose reduction while further improving image quality in CCTA for overweight patients.
More Related Videos
11:09High-Resolution Cardiac Positron Emission Tomography/Computed Tomography for Small Animals
Published on: December 16, 2022
09:32Time-Resolved, Dynamic Computed Tomography Angiography for Characterization of Aortic Endoleaks and Treatment Guidance via 2D-3D Fusion-Imaging
Published on: December 9, 2021
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Radiological Investigation I: X-ray and CT