Validation of a Remote Monitoring Blood Pressure Device in Pregnancy
Ayamo Oben1,2, Matthew Moore2, Eric Wallace2,3
1Department of Obstetrics and Gynecology, University of Alabama at Birmingham, Birmingham, Alabama, USA.
Background:
The Ideal Life Blood Pressure Manager measures blood pressure (BP) and automatically transmits results to the patient's medical record independent of internet access, but has not been validated. Our objective was to conduct a validation study of the Ideal Life BP Manager in pregnant women using a validation protocol.
Methods:
Pregnant participants were enrolled into three subgroups per the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization protocol: normotensive (systolic blood pressure (SBP) < 140 and diastolic blood pressure (DBP) < 90), hypertensive without proteinuria (SBP ≥ 140 or DBP ≥ 90), and preeclampsia (SBP ≥ 140 or DBP ≥ 90 with proteinuria). Two trained research staff used a mercury sphygmomanometer to validate the device, alternating sphygmomanometer, and device readings for a total of 9 measurements.
Results:
Among 51 participants, the mean SBP and DBP differences and standard deviations between the device and the mean staff measurements for all participants were 1.7 ± 7.1 and 1.5 ± 7.0 mm Hg, respectively. The standard deviations of the individual participant's paired device and mean staff SBP and DBP measurements were 6.0 and 6.4 mm Hg, respectively. The device was more likely to overestimate rather than underestimate BP (SBP: mean difference = 1.67, 95% CI [-12.15 to 15.49]; DBP: mean difference = 1.51, 95% CI [-12.26 to 15.28]). Most paired readings had a difference of less than 10 mm Hg across averaged paired readings.
Conclusion:
The Ideal Life BP Manager met internationally recognized validity criteria in this sample of pregnant women.
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Errors occurring during blood pressure monitoring
Several factors...
Pre-Procedural Guidelines for Assessing Blood Pressure
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment: