Related Experiment Video
Updated: Aug 9, 2025

Fully Endoscopic Mitral Valve Repair with Percutaneous Cannulation of Groin Vessels
Published on: May 26, 2023
Analysis of Reported Adverse Events Associated with Over-the-Scope Endoscopic Suturing System: an FDA MAUDE Database
Natalie Wilson1, Mohamed Abdallah2, Fouad Jaber3
1Department of Internal Medicine, University of Minnesota, Minneapolis, MN, 55455, USA.
Purpose:
The over-the-scope endoscopic suturing system (ESS) (OverStitch™) is one of the most widely utilized endoscopic suturing systems in current clinical practice; however, data on the adverse events associated with this device is scarce. Our study aims to evaluate the adverse events and complications associated with the over-the-scope ESS using the FDA's Manufacturer and User Facility Device Experience (MAUDE) database.
Materials And Methods:
We analyzed the post-marketing surveillance data from the FDA MAUDE database for the over-the-scope ESS from January 2008 through June 2022.
Results:
Eighty-three reports were filed from January 2008 to June 2022. Adverse events were classified as device-related complications and patient-related adverse events. Seventy-seven device-related issues and 87 patient adverse events were identified. The most common device-related issue was difficulty to remove after deployment (n=12, 15.58%) followed by mechanical problem (n=10, 12.99%), mechanical jam (n=9, 11.69%), or entrapment of device (n=9, 11.69%). Of the 87 patient-related adverse events, the most common was perforation (n=19; 21.84%), followed by device embedded in tissue or plaque (n=10; 11.49%), and abdominal pain (n=8; 9.20%). Of the 19 patients who experienced perforation, two required open surgical repair and one required laparoscopic surgical repair.
Conclusion:
The overall adverse events from the over-the-scope ESS remain acceptable as evidenced by the number of reported cases since 2008. However, it is important to note that adverse event rates might increase as the use of the device increases; therefore, it is essential for endoscopists to be aware of the potential common and rare adverse events associated with the use of the over-the-scope ESS device.
More Related Videos
10:52Laparoscopic Repair of Para-Esophageal Hernia Using Absorbable Biosynthetic Mesh
Published on: September 11, 2021
03:24Author Spotlight: Development and Application of a Novel Suture Technique for Annular Fibrosus Repair in Percutaneous Transforaminal Endoscopic Discectomy
Published on: January 26, 2024
Related Concept Videos
Pharmacovigilance
This process, termed pharmacovigilance, aims to detect, evaluate, and minimize harmful effects related to medication use. The data collection for pharmacovigilance depends on spontaneous reporting systems, where healthcare professionals or patients voluntarily report suspected ADRs.
In some cases, there...
Endoscopic Procedures I: Esophagogastroduodenoscopy
During an EGD, the endoscope can be used to: