Data collection and analysis from a respiratory inductance plethysmograph.
N B Hodsman1, G N Kenny, R Rennie
1University Department of Anaesthesia, Glasgow Royal Infirmary.
Summary
Respitrace, a respiratory inductance plethysmograph, accurately measured respiratory rates in patients receiving narcotic analgesia post-surgery. However, frequent nursing interventions impacted tidal volume and phase angle accuracy.
Area of Science:
- Anesthesiology
- Respiratory Physiology
- Medical Instrumentation
Background:
- Narcotic analgesia is commonly used for postoperative pain management.
- Understanding the respiratory effects of these medications is crucial for patient safety.
- Respiratory inductance plethysmography offers a non-invasive method to monitor breathing patterns.
Purpose of the Study:
- To evaluate the respiratory effects of narcotic analgesia in the first 24 hours after surgery.
- To assess the accuracy of Respitrace in measuring respiratory parameters during this period.
- To identify factors influencing the reliability of respiratory monitoring.
Main Methods:
- Utilized a respiratory inductance plethysmograph (Respitrace) for continuous monitoring.
- Collected data online using an Apple microcomputer for analysis of respiratory rates and tidal volumes.
- Analyzed central and obstructive apnea using the phase angle between abdominal and respiratory components.
- Employed data compression techniques to manage large datasets (over 20,000 breaths/patient).
Main Results:
- Respiratory rates were accurately measured breath-by-breath.
- Tidal volume and phase angle measurements were less accurate due to frequent nursing interventions.
- The Respitrace system successfully collected extensive respiratory data.
Conclusions:
- Respitrace can accurately monitor respiratory rates in postoperative patients on narcotic analgesia.
- Nursing interventions can compromise the accuracy of tidal volume and phase angle measurements.
- Further refinement of data collection protocols may be needed to improve accuracy in clinical settings.