Related Experiment Video
Updated: Aug 9, 2025

Absolute Quantification of Aβ1-42 in CSF Using a Mass Spectrometric Reference Measurement Procedure
Published on: March 21, 2017
Development of a candidate reference measurement procedure by ID-LC-MS/MS for total tau protein measurement in
Chiara Giangrande1, Hélène Vaneeckhoutte1, Amandine Boeuf1
1Department of Biomedical and Organic Chemistry, Laboratoire National de Métrologie et d'Essais (LNE), Paris, France.
Objectives:
In clinical pratice, tau protein measurement generally relies on immunoassays (IAs), whose major drawback is the lack of results comparability due to differences in selectivity and/or calibration. This underlines the importance of establishing a traceability chain for total tau (t-tau) measurements. The objective of this work is to develop a higher order candidate reference measurement procedure (RMP) for the absolute quantification of t-tau in cerebrospinal fluid (CSF).
Methods:
To calibrate the candidate RMP and establish metrological traceability to the SI units, a primary calibrator consisting in a highly purified recombinant protein was sourced. Its purity was evaluated by liquid chromatography coupled with high resolution mass spectrometry (LC-HRMS) and the protein mass fraction in solution was certified by amino acid analysis (AAA). An isotopically-labelled homologue was obtained to develop a candidate RMP by isotope dilution mass spectrometry (IDMS) for t-tau absolute quantification in CSF. Calibration blends and quality control (QC) materials were gravimetrically prepared and subjected to the same preparation workflow as CSF samples, followed by LC-HRMS analysis in Parallel Reaction Monitoring (PRM) mode.
Results:
A primary calibrator has been developed and an IDMS candidate RMP has been validated for CSF t-tau. The candidate RMP was used to certify t-tau concentration in three pools of CSF (low, medium, high).
Conclusions:
The candidate RMP will pave the road towards global standardization of CSF t-tau measurements. Together with commutable Certified Reference Materials (CRMs), it will allow evaluating and improving the accuracy and comparability of results provided by IAs.

