Undefilled blood tube containing EDTA: Is it an inappropriate sample for HbA1c assay?
Berrak Guven1, Ismail Benice1, Murat Can1
1Department of Biochemistry, Faculty of Medicine, Zonguldak Bülent Ecevit University, Zonguldak, Turkey.
Introduction:
Blood samples having inappropriate volume are a substantial part of preanalytical errors. Inadequate sample volume for glycated haemoglobin (HbA1c) test may be a common problem of patients with diabetes mellitus having vascular changes. In this study, we compared HbA1c concentrations of underfilled and appropriately filled blood collection tubes.
Materials And Methods:
To compare HbA1c concentrations, blood samples were collected into 2 mL tubes containing K3-EDTA from 109 subjects. Two blood samples (underfilled and appropriately filled) were drawn from a patient by the same personnel and materials. HbA1c measurements were assayed on a Cobas 6000 analyser module c 501 (Roche Diagnostics, Mannheim, Germany). The HbA1c% results were compared by t-test and Wilcoxon's signed-rank statistical methods (SPSS Inc., Chicago, USA). Bias analysis was performed using Microsoft Excel 4.0.
Results:
Underfilled samples were classified three groups (group 1, N = 44; group 2, N = 36; and group 3, N = 29) according to the filling ratio of the samples; 0.5 mL and below (< 25%), 0.5-1.0 mL (25-50%), and 1.0-2.0 mL (> 50%), respectively. When we compared underfilled tubes with pairing filled tubes, there was a statistically significant difference only with tubes filled less than 25% (P = 0.030). Furthermore, we have done bias analysis between paired tubes according to the diagnostic cut-off value of 6.5%. The bias was more prominent in up to 50% underfilled blood tubes (1.1%), when HbA1c concentrations were below the diagnostic cut-off of 6.5%.
Conclusions:
We suggest that the blood tubes with EDTA for HbA1c measurement should be filled with at least 50% to avoid clinical variations.
More Related Videos
08:44An In Vitro Hemodynamic Loop Model to Investigate the Hemocytocompatibility and Host Cell Activation of Vascular Medical Devices
Published on: August 21, 2020
07:24A Precision Medicine Tool for Measurement and Monitoring of Hemoglobin S in Sickle Cell Disease Patients Receiving Transfusion Therapy
Related Concept Videos
Effects of EDTA on End-Point Detection Methods
In the visual method, metal-ion indicators (metallochromic dyes), which have distinct colors in their free and complex forms, are added to the mixture to signal the titration's end point. They form stable complexes with metal ions, but these complexes are weaker than the corresponding metal–EDTA complexes. As a...
Complexometric EDTA Titration Curves
EDTA: Indirect and Alkalimetric Titration
EDTA: Auxiliary Complexing Reagents
EDTA: Chemistry and Properties
EDTA: Direct, Back-, and Displacement Titration
Direct titration involves buffering the metal ion solution to the desired pH and directly titrating with standard EDTA until the endpoint. The optimum pH ensures a large conditional formation constant of metal−EDTA and visibility of the free indicator color in the solution. In addition, auxiliary complexing reagents are used to prevent the precipitation of metal hydroxides...