Cortical Bone Assessment Using Ultrasonic Guided Waves: A Reproducibility Study in a Healthy Population
Published on: January 31, 2025
The new fast kilovoltage-switching dual-energy computed tomography for measuring bone mineral density
Mingyue Wang1, Yan Wu1, Yue Zhou1
1Department of Radiology, The First Affiliated Hospital of Zhengzhou University, Zhengzhou, China.
Background:
The update in technology may impact the accuracy in measuring bone mineral density (BMD). However, the application of the new fast kilovoltage (kV)-switching dual-energy computed tomography (DECT) for BMD measurement has not yet been reported. This study aimed to examine the accuracy and precision of the new fast kV-switching DECT in measuring BMD and to evaluate its applicability in clinical BMD measurement.
Methods:
Forty sets of the new fast kV-switching DECT scans and one quantitative computed tomography (QCT) scan were performed on the European Spine Phantom. Their relative errors and relative standard deviations were compared. A retrospective analysis was performed on patients who underwent chest plain DECT and abdominal monoenergetic plain CT at the same time. The relationship between hydroxyapatite-water and hydroxyapatite-fat measured using DECT and BMD measured using QCT was analyzed by multivariate regression analysis.
Results:
The relative errors of the new fast kV-switching DECT with low tube speeds (0.8 and 1.0 s/r) were all less than 6% and were less than those of QCT, except for those at 515 mA. The relative standard deviation values with high tube rotation speeds (0.5 and 0.6 s/r) were higher than those with low tube speeds (0.8 and 1.0 s/r) under most tube current conditions. The new fast kV-switching DECT-derived BMD values corrected by multiple linear regression (predicted hydroxyapatite) were significantly positively correlated with the QCT-based BMD values (R2=0.912; P<0.001). The results of the Bland-Altman analysis demonstrated high consistency between the 2 measurement methods.
Conclusions:
Results of the phantom measurements indicated that the new fast kV-switching DECT could measure BMD with relatively high accuracy and precision. The results of a subsequent clinical in vivo experiment demonstrated that vertebral BMD measurements derived from DECT and QCT were mostly consistent and highly accurate. Therefore, patients who undergo DECT for other clinical indications can simultaneously have their BMD determined.