Effect of a Motion Artifact Correction System on Cone-Beam Computed Tomography Image Characteristics
Mayank Pahadia1, Rujuta Katkar2, Hassem Geha3
1Oral and Maxillofacial Diagnostic Sciences, University of Florida, Gainesville, USA.
Objectives:
Determine the effect of the motion correction system on cone-beam computed tomography (CBCT) image quality parameters, artifacts, and contrast-to-noise ratio (CNR) using different motion settings.
Materials And Methods:
A customized phantom insert array was prepared using SEDENTEX CT IQ Phantom (Leeds Test Objects, Yorkshire, England) stabilized over a rotating electric turntable. Thirty baseline CBCT scans were acquired with standardized technique factors on the ProMax 3D (Planmeca, Helsinki, Finland) machine using combinations of different motion settings, including no motion, three- and six-degree motion, and with and without the use of a motion correction system. The standardized images were exported to ImageJ software. Image quality parameters, artifacts, and CNR values were evaluated and compared among the different acquisition settings.
Results:
The use of the motion correction system algorithm compared with the different motion settings showed a statistically significant difference for all the parameters (p<0.05) except for artifact values for six-degree motion (p<0.07). The effect of different motion settings on the parameters was not statistically significant.
Conclusion:
The use of a motion correction system, a proprietary algorithm-based system incorporated in the ProMax 3D CBCT unit, deteriorates the image quality characteristics evaluated in this in vitro study, namely artifact value and CNR. Its use in clinical settings might be limited to situations where patient motion is expected and appropriate head stabilization is not possible due to age or disease.
More Related Videos
06:53Management of Respiratory Motion Artefacts in 18F-fluorodeoxyglucose Positron Emission Tomography using an Amplitude-Based Optimal Respiratory Gating Algorithm
Published on: July 23, 2020
10:23Author Spotlight: Three-Dimensional Cephalometric Landmark Annotation Demonstration on Human Cone Beam Computed Tomography Scans
Published on: September 8, 2023
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography